Polyethylene particles in synovial fluid after knee arthroplasty with a conventional or highly cross-linked polyethylene. Preliminary study
Partículas de polietileno en líquido sinovial tras artroplastia de rodilla con un polietileno convencional o uno altamente entrecruzado. Estudio preliminar
Visits
1844
P. Hinarejosa,
Corresponding author
Phinarejos@parcdesalutmar.cat

Corresponding author.
, A. Torresa, I. Piñola, D. Valverdea, E. Pratsb, Ll. Puiga
a Servicio de Cirugía Ortopédica y Traumatología, Parc de Salut Mar (Hospitales del Mar y la Esperanza), Barcelona, Spain
b Servicios Científico-Técnicos de la Universidad de Barcelona, Departamento de Microscopía Electrónica de Barrido, Barcelona, Spain
This item has received
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Tables (4)
Table 1. Preoperative variables analysed in both groups.
Table 2. Implant variables analysed in both groups.
Table 3. Postoperative variables analysed in both groups.
Table 4. Results of the electronic microscope study on synovial fluid samples.
Show moreShow less
Abstract
Aim of the study

In recent years cross-linked polyethylenes have been developed in an attempt to reduce the wear, as has been demonstrated in knee simulators. The aim is to assess, by counting particles of polyethylene in synovial fluid, whether the reduction in wear is confirmed in patients with a highly crosslinked polyethylene prosthesis.

Material and methods

A prospective randomised study was designed. During the implantation of a knee prosthesis, one group of patients was assigned the use of a conventional polyethylene (group A), and the other group a highly crosslinked polyethylene (X3®, Stryker Orthopaedics) (group B). At 12 months after surgery a knee arthrocentesis was performed, and the number of polyethylene particles was counted in a scanning electron microscopy. Fourteen samples from each group were studied.

Results

Both groups were comparable in all study variables. We found no significant differences in the concentration of polyethylene particles/ml (1.49±0.85 million in group A vs 1.42±0.91 million in group B, P=0.60) or the total number of isolated particles. We found no differences either in size or morphology of particles between both groups.

Discussion and conclusions

Although several in vitro studies in vitro using different types of highly crosslinked polyethylene found a significant reduction, we did not find that that wear was reduced in the knees of these patients. The great variability in the number of particles between individuals suggests that polyethylene wear in vivo depends on many factors, so perhaps the type of polyethylene is not the most significant factor.

Keywords:
Total knee replacement
Polyethylene
Particles
Synovial fluid
Resumen
Objetivo

En los últimos años han aparecido polietilenos altamente entrecruzados para intentar disminuir el desgaste, tal como se ha demostrado en simuladores de rodilla. El objetivo es evaluar mediante el recuento de partículas de polietileno en líquido sinovial, si se confirma un menor desgaste en pacientes con prótesis de polietileno altamente entrecruzado.

Material y método

Estudio prospectivo aleatorizado, en el que durante la implantación de una prótesis de rodilla se asignó a un grupo de pacientes la colocación de un polietileno convencional (grupo A), y a otro grupo un polietileno altamente entrecruzado (X3® de Stryker Orthopaedics) (grupo B). A los 12 meses tras la cirugía se practicó una artrocentesis de rodilla y se hizo un recuento de partículas de polietileno en el líquido sinovial mediante microscopio electrónico de barrido. Se han analizado 14 muestras en cada grupo.

Resultados

Ambos grupos son comparables en todas las variables estudiadas. No hemos hallado diferencias significativas en la concentración de partículas de polietileno/ml (1,49±0,85 millones grupo A vs. 1,42±0,91 millones grupo B; p=0,60) ni en el número total de partículas aisladas entre ambos grupos. Tampoco hemos hallado diferencias en el tamaño ni la morfología de partículas entre ambos grupos.

Discusión y conclusiones

Aunque diversos trabajos in vitro han hallado una reducción muy significativa del desgaste del polietileno altamente entrecruzado, no se ha hallado que en las prótesis de rodilla implantadas en pacientes este desgaste se vea reducido. La gran variabilidad del número de partículas entre individuos sugiere que el desgaste de polietileno in vivo depende de muchos factores y probablemente el tipo de polietileno no sea el más determinante.

Palabras clave:
Prótesis total de rodilla
Polietileno
Partículas
Líquido sinovial
Full Text
Introduction

Osteolysis is one of the main causes of failure in total knee arthroplasty. Consequently, osteolysis secondary to polyethylene wear and aseptic loosening is the principle cause of revision starting from 5 years after the index operation.1,2

In knee arthroplasty, the chief wear mechanism is fatigue. It produces larger particles than those produced in hip arthroplasty, in which wear occurs from abrasion and adhesion.3 The size of the particles is important, given that the smaller ones have greater biological activity and producer greater osteolysis.3 Particles >10μm are not very pro-inflammatory because they are not susceptible to phagocytosis.4

Osteolysis is the result of a foreign-body reaction, induced by the particles generated by polyethylene wear.5 These particles are phagocytosed by macrophages and giant cells, which are activated and liberate cytokines. Interleukins IL-1β, IL-6 and tumour necrosis factor alpha produce inflammation and stimulate the osteoclast; the activated osteoclast is the cell responsible for the osteolysis6–8; however, the cytokines also inhibit bone formation by the osteoblasts.9

To decrease polyethylene particle release, the industry has introduced different improvements over the last decades: improving the polyethylene sterilisation processes, using gamma radiation in an oxygen-free medium,10 introducing new friction pairs and developing highly cross-linked polyethylene (XLPE).11,12 Among these XLPE is the X3® (Stryker Orthopaedics, Mahwah, NJ).

The XLPE have been used for over 10 years in total hip arthroplasty, where they have shown better in vivo wear results than conventional polyethylene.13–15 However, the studies carried on knee arthroplasty, and in those based on current use in this joint, are in vitro studies on a simulator.16,17 In one study, Wang6 found a reduction in polyethylene wear of 68% and 64% in standard arthroplasties with posterior cruciate ligament retention (CR) and posterior-stabilised (PS) total knee prosthesis (TKP) respectively, comparing polyethylene X3® to conventional polyethylene.1 However, we know that simulator and in vitro studies do not always faithfully reproduce the conditions found in clinical practice.

The hypothesis of this study was that using XLPE X3® would reduce the number of polyethylene particles from wear, compared to conventional polyethylene conventional in an in vivo context. The main study objective was to compare the number of polyethylene particles isolated in synovial fluid following primary total knee arthroplasty, as based on using a conventional or XLPE polyethylene. The secondary objective was comparing particle size and morphology in both groups.

Material and methods

We designed a random, prospective study. The study protocol was approved by the local Ethics Committee. Patients gave informed consent for their inclusion in the study, which covered synovial fluid extraction by arthrocentesis at 12 months.

In all of the cases, the implant was the Triathlon® TKP posterior stabilised (PS) bone cement type (Stryker Orthopaedics, Kalamazoo, MI, USA), with the same surgeon performing all of the operations.

The patients were randomised by intraoperative closed-enveloped assignment, putting 50% of the patients in group A (conventional polyethylene) and 50% in group B (XLPE, or X3®). The conventional polyethylene was high molecular weight polyethylene (GUR 1020) irradiated at 30kGy (gamma irradiation) in nitrogen. The X3® polyethylene was high molecular weight XLPE, submitted to 3 irradiation cycles at 30kGy in air, followed by air annelation at 130°C for 8hours. After the 3 cycles, it was sterilised with gas-plasma.

The criteria for inclusion were the implantation of a primary PS TKP, in patients aged between 65 and 85 years.

The criteria for exclusion were a diagnosis other than primary gonarthrosis (post-traumatic, inflammatory arthropathy), the presence of severe preoperative deformity (>12° of varus or valgus in the tibiofemoral axis), the presence of postoperative varus or valgus alignment >4°, voluntary withdrawal from the study and the impossibility of obtaining 3cc of synovial fluid at the 12-month control visit.

The variables considered in the preoperative assessment were age, sex, weight, height, body-mass index (BMI), tibiofemoral angle (TFA) measured in telemetric radiological projection, and the knee and function values using the Knee Society Score (KSS) scale.18

The mean patient age was 73.1±6.2 years; there were 8 males and 20 females. The preoperative KSS scale score was 42.6±13.6 points in the knee subscale and 55.4±7.8 for function.

The implant variables considered were size of the femoral and tibial components and polyethylene thickness. The most commonly used femoral and tibial sizes used were 3 and 4. The polyethylene used was the thinnest, 9mm thick in 82% of the cases.

The patients had the normal follow-ups in the postoperative period. In the assessment at 12 months after the operation, the following variables were considered: KSS scale knee and function values, postoperative TFA measured in telemetry radiology projection and the result of the physical activity questionnaire according to the UCLA scale.19 The 12-month KSS scale values were 92.4±3.8 points on the knee subscale and 80.9±16.3 for function. The mean postoperative TFA was 0.9±1.9°. In the UCLA scale, 57% of the patients were in Category 6, 21% in Category 7, 14% in Category in Category 5 and 7% in Category 4.

All of the patients received an arthrocentesis under sterile conditions at the 12-month follow-up, with the same technician having evacuated the maximum amount of synovial fluid. In all cases, the minimum volume was 3ml; the mean volume obtained was 12.4±6.9ml.

Once the synovial fluid samples were identified, they were stored in a sterile tube in a freezer at a −20°C until they were processed in the following weeks. The samples were then processed according to the technique described by Minoda et al.,20 which basically consisted of digesting the sample by sodium hydroxide at 65°C for 12hours to destroy the tissue residues. Next, a continuous sucrose slope was applied in a 14-ml tube and ultracentrifuged at 28,000rpm for 3hours at 4°C. The top layer of the tube was immediately taken and an isopropanol–water slope was applied in a 40-ml tube likewise ultracentrifuged at 28,000rpm for an hour at room temperature to group the particles. Finally, the inter-stage line that was filtered through a 0.1-μm polycarbonate filter was taken, which was where the polyethylene particles were trapped, placing the filter in a Petri dish.

In the scanning electron microscopy service, the filter was mounted on a brass stub with conducting bioadhesive tabs and colloidal silver; this was then covered with a 40-nm gold stub and analysed with the microscope. The microscope used was a Quanta 200 field emission gun (FEG) environmental scanning electron microscope (ESEM) (FEI, Co, USA). Using this instrument, each sample was analysed in 10 random fields at 10,000× by a specialised technician. In cases of doubt as to the particle composition, the sample was analysed with an energy-dispersive microanalysis system (EDAX Genesis and Polaron E5000 cathode sputter diode); this made it possible for us to know whether what we had was really a polyethylene particle or a contaminant particle.

In each high magnification field, the number of particles, their means and their morphology were analysed.

Statistical analysis

We used the statistical package SPSS 15.0 for Windows for the statistical analysis. Quantitative variables were compared with the Mann–Whitney U-test and categorical variables were analysed with the Chi square or Fisher exact test, as applicable. Statistical significance was set at P<0.05.

Results

Both groups were comparable in all the preoperative variables, including the anthropometric and radiological variables and the KSS score (Table 1). Likewise, there were no significant differences in implant size or polyethylene thickness (Table 2). The 2 groups were also comparable in the postoperative variables: TFA, UCLA scale and KSS scale score (Table 3).

Table 1.

Preoperative variables analysed in both groups.

  Conventional  X3®  P 
Age  73.6±4.9  72.6±7.4  0.769 
Gender  4 M/10 F  4 M/10 F 
Weight  70.7±13.2  74.0±12.2  0.375 
Height  155.0±9.2  156.3±8.6  0.595 
BMI  29.9±6.3  30.2±3.7  0.494 
KSS-Knee  43.1±15.6  41.9±11.6  0.769 
KSS-Function  54.3±7.6  56.4±8.1  0.511 
Table 2.

Implant variables analysed in both groups.

  Conventional  X3®  P 
Femur size  3.5±1.0  3.9±1.1  0.427 
Tibia size  3.6±1.1  4.1±1.1  0.285 
Polyethylene thickness  9.3±0.7  9.6±1.2  0.734 
Table 3.

Postoperative variables analysed in both groups.

  Conventional  X3®  P 
P.O. TF angle  −0.6±1.9  −1.2±2.2  0.511 
UCLA  5.9±0.7  5.9±0.9  0.982 
P.O. KSS-Knee  93.3±2.9  92.1±4.8  0.946 
P.O. KSS-Function  78.6±14.8  83.6±16.5  0.310 

The volume of the arthrocentesis-obtained synovial fluid was 11.7±7.2ml in group A (conventional polyethylene) and 13.1±5.7ml in group B (X3®polyethylene) (P=0.635).

The mean polyethylene particle concentration/ml was 1.46×106±0.88×106. The total number of polyethylene particles per sample was 18.63×106±17.93×106. We found no differences in polyethylene particle concentration/ml or in the total number of particles isolated between the samples in groups A and B (Table 4).

Table 4.

Results of the electronic microscope study on synovial fluid samples.

  Conventional  X3®  P 
Particles/ml (millions)  1.488±0.848  1.423±0.912  0.603 
Total particles (millions)  17.246±16.929  20.059±18.995  0.701 
% of particles >1μm  2.38±3.0  4.43±3.8  0.104 
% of non-round particles  24.8±10.5  21.3±12.2  0.246 

The polyethylene particles isolated varied greatly in size, from 0.1μm to 30μm, but 96.57% were equal to or less than 1μm in diameter. Comparing the percent of particles larger than 1μm between both groups revealed a tendency towards a greater percentage of larger particles in group B, although the difference (P=0.104) was not statistically significant (Table 4).

With respect to particle morphology, we found 76.5% of them to be round. There was no difference in the percentage of non-round particles between the 2 groups (Table 4).

Discussion

The main study finding was the lack of difference in the number of polyethylene particles in synovial fluid between the group with the knee prosthesis made of conventional polyethylene and the group having ones made of XLPE. There are few articles in the literature on the study of polyethylene particles in synovial fluid in patients and, as far as we know, only one compared 2 groups with different types of polyethylene. However, that was a retrospective study, analysing few cases.21

The number of particles isolated in that study (some 18.7 million per sample) compared positively with other series using metal–polyethylene TKP (57 million). However, the number compared negatively with respect to TKP with frictional torque ceramic–polyethylene (7.1 million).22

Various studies have compared polyethylene wear in vitro previously between an XLPE polyethylene and a conventional one in a knee simulator subjected to several million cycles. Poppola23 found a 72–85% wear reduction with NexGen® XLPE (Zimmer, Warsaw, Ind, USA). Other in vitro studies using Stryker's X3® XLPE also found reductions in wear above 60%.16,24

There are hardly any studies analysing XLPE wear in vivo. In one TKP study, the authors found reduced wear analysing the particles in synovial fluid, but these results were based on only 7 cases (3 in 2 group and 4 in the other).21 In contrast, in a study on hip prostheses revised, the XLPE implants extracted showed wear similar to that of the conventional polyethylene.25

In our study, we found a reduction of only 4.37% in particles/ml of synovial fluid, which is not statistically significant. In addition, there was a great variability among individuals, suggesting that in vivo, apart from the type of polyethylene, many other factors affect wear: patient weight and activity, the three-dimensional alignment of the components, the size of the prosthetic components (which determines its surface), ligament tension and many others.

The adverse biological reactions that provoke osteolysis do not depend on only wear factors; they also depend on particle size and concentration.26 In reference to particle size, the first generation of XLPE generated larger particles, which has been associated with increased capability to generate osteolysis.27 Nevertheless, other studies using contemporary XLPE revealed no differences in particle size in comparison to conventional polyethylene,28,29 even if the friction surface was ceramic instead of metal.30 In this study, we did not find any significant differences in particle size between the 2 groups either. However, we did observe that the percentage of particles larger than 1μm tended to be greater in the XLPE group.

Likewise, although there is evidence that elongated particles are more pro-inflammatory than round ones,4 we did not find significant differences in the polyethylene particle morphology between the 2 groups.

There has been speculation that the post in the SP component might be more fragile.31 However, from a clinical point of view, a few studies have shown that using XLPE in knee prostheses is safe in the short and medium term, but that it does not provide any advantages with respect to clinical or radiological results.32,33

Our study has a few limitations, such as the small number of cases and how difficult it is to control some variables that can affect polyethylene wear as much as or more than the type itself (for example, ligament tension, which is impossible to measure in vivo). There is also the fact that we used 0.1-μm pore filters as in the technique described by Minoda20 to process the synovial fluid processing; consequently, the particles smaller than this were not trapped in the filter. However, previous studies have shown that most of the TKP wear particles fall in the range between 0.1 and 1μm34,35 and the particles that provoke the greatest osteolysis are those that measure from 0.3 to 1μm.36 Therefore, the particles not retained by the filter that we used would hardly be causes of osteolysis.

Another limitation in the study is the follow-up length, limited to 12 months. Even though we chose the same figure used in other previous studies that compared different TKP,22,37 we cannot rule out that the lack of differences would be modified as the polyethylene aged, as some in vitro studies have demonstrated.31,38

In conclusion, we can state that no in vivo reduction in the number of polyethylene particles was found at 12 months using XLPE. The great variability observed between individuals suggests that polyethylene wear is multifactorial and that the type of polyethylene is probably not the main factor that affects it.

Level of evidence

Level of evidence I.

Ethical Disclosures

Protection of human and animal subjects. The authors will declare that the procedures followed were in accordance with the regulations of the responsible Clinical Research Ethics Committee and in accordance with those of the World Medical Association and the Helsinki Declaration.

Confidentiality of Data. The authors declare that no patient data appears in this article.

Right to privacy and informed consent. The authors declare that no patient data appears in this article.

Funding

This study was funded by a fellowship awarded by the SECOT Foundation in the 2009 application process.

Conflict of interests

The authors have no conflict of interests to declare.

Acknowledgements

Thanks are expressed to the SECOT Foundation for the 2009 fellowship for the preparation of this study, to Gabriel Gil at the Institut Municipal d’Investigació (IMIM) in Barcelona for his collaboration in the sample processing, and to Sergi Mojal (IMIM) for his collaboration in the statistical study.

References
[1]
T.K. Fehring, S. Odum, W.L. Griffin, J.B. Mason, M. Nadaud.
Early failures in total knee arthroplasty.
Clin Orthop Relat Res, 392 (2001), pp. 315-318
[2]
P.F. Sharkey, W.J. Hozak, R.H. Rothman, S. Shastri, S.M. Jacoby.
Insall Award paper. Why are total knee arthroplasty failing today?.
Clin Orthop Relat Res, 404 (2002), pp. 7-13
[3]
T.P. Schmalzried, M. Jasty, A. Rosenberg, W.H. Harris.
Polyethylene wear debris and tissue reactions in knee as compared to hip replacement prostheses.
J Appl Biomater, 5 (1994), pp. 185-190
[4]
N.J. Hallab, J.J. Jacobs.
Biologic effects of implant debris.
Bull NYU Hosp Jt Dis, 67 (2009), pp. 182-188
[5]
R.M. Baxter, T.A. Freeman, S.M. Kurtz, M.J. Steinbeck.
Do tissues from THA revision of highly crosslinked UHMWPE liners contain wear debris and associated inflamation.
Clin Orthop Relat Res, 469 (2011), pp. 2308-2317
[6]
J. Chiba, L.J. Schwenderman, R.E. Booth Jr., L.S. Crossett, H.E. Rubash.
A biochemical, histologic, and immunohistologic analysis of membranes obtained from failed cemented and cementless total knee arthroplasty.
Clin Orthop Relat Res, 299 (1994), pp. 114-124
[7]
E. Ingham, J. Fisher.
The role of macrophages in osteolysis of total joint replacement.
Biomaterials, 26 (2005), pp. 1271-1286
[8]
J.J. Jacobs, K.A. Roebuck, M. Archibeck, N.J. Hallab, T.T. Glant.
Osteolysis: basic science.
Clin Orthop Relat Res, 393 (2001), pp. 71-77
[9]
S.K. Gupta, A. Chu, A.S. Ranawat, J. Slamin, C.S. Ranawat.
Osteolysis after total knee arthroplasty.
J Arthroplasty, 22 (2007), pp. 787-799
[10]
A.A. Besong, J.L. Tipper, E. Ingham, M.H. Stone, B.M. Wroblewski, J. Fisher.
Quantitative comparison of wear debris from UHMWPE that has and has not been sterilised by gamma irradiation.
J Bone Joint Surg Br, 80 (1998), pp. 340-344
[11]
R.D. Crowninshield, O.K. Muratoglu.
Implant Wear Symposium 2007 Engineering Work Group. How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement.
J Am Acad Orthop Surg, 16 (2008), pp. S80-S85
[12]
E. Oral, O.K. Muratoglu.
Vitamin E diffused, highly crosslinked UHMWPE: a review.
Int Orthop, 35 (2011), pp. 215-223
[13]
L.D. Dorr, Z. Wan, C. Shahrdar, L. Sirianni, M. Boutary, A. Yun.
Clinical performance of a Durasul Highly cross-linked polyetylene acetabular liner for total hip arthroplasty at five years.
J Bone Joint Surg Am, 87 (2005), pp. 1816-1821
[14]
D.W. Manning, P.P. Chiang, J.M. Martell, J.O. Galante, W.H. Harris.
In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty.
J Arthroplasty, 20 (2005), pp. 880-886
[15]
S.M. Kurtz, H.A. Gawel, J.D. Patel.
History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene.
Clin Orthop Relat Res, 469 (2011), pp. 2262-2277
[16]
A. Wang, S. Yau, A. Essner, L. Herrera, M. Manley, J. Dumbleton.
A highly crosslinked UHMWPE for CR and PS total knee arthroplasty.
J Arthroplasty, 23 (2008), pp. 559-566
[17]
P.A. Williams, C.M. Brown, R. Tsukamoto, I.C. Clarke.
Polyethylene wear debris produced in a knee simulator model: effect of crosslinking and counterface material.
J Biomed Mater Res B: Appl Biomater, 92 (2010), pp. 78-85
[18]
J.N. Insall, L.D. Dorr, R.D. Scott, W.N. Scott.
Rationale of the Knee Society clinical rating system.
Clin Orthop Relat Res, 248 (1989), pp. 13-14
[19]
C.A. Zahiri, T.P. Schmalzried, E.S. Szuszczewicz, H.C. Amstutz.
Assessing activity in joint replacement patients.
J Arthroplasty, 13 (1998), pp. 890-895
[20]
Y. Minoda, A. Kobayashi, H. Iwaki, M. Miyaguchi, Y. Kadoya, H. Ohashi, et al.
Polyethylene wear particles in synovial fluid after total knee arthroplasty.
Clin Orthop Relat Res, 410 (2003), pp. 165-172
[21]
K. Iwakiri, Y. Minoda, A. Kobayashi, R. Sugama, H. Iwaki, F. Inori, et al.
In vivo comparison of wear particles between highly crosslinked polyethylene and conventional polyethylene in the same design of total knee arthroplasties.
J Biomed Mater Res B: Appl Biomater, 91 (2009), pp. 799-804
[22]
Y. Minoda, A. Kobayashi, H. Iwaki, M. Miyaguchi, Y. Kadoya, H. Ohashi, et al.
Characteristics of polyethylene particles isolated from sinovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties.
J Biomed Mater Res B: Appl Biomater, 71 (2004), pp. 1-6
[23]
O.O. Popoola, J.Q. Yao, T.S. Johnson, C.R. Blanchard.
Wear, delamination, and fatigue resistance of melt-annealed highly crosslinked UHMWPE cruciate-retaining knee inserts under activities of daily living.
J Orthop Res, 28 (2010), pp. 1120-1126
[24]
M.A. Kester, L. Herrera, A. Wang, A. Essner.
Knee bearing technology. Where is technology taking us?.
J Arthroplasty, 22 (2007), pp. S16-S20
[25]
D.T. Schroder, N.H. Kelly, T.M. Wright, M.L. Parks.
Retrieved hightly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE.
Clin Orthop Relat Res, 469 (2011), pp. 387-394
[26]
J. Fisher, J. Bell, P.S. Barbour, J.L. Tipper, J.B. Matthews, A.A. Besong, et al.
A novel method for the prediction of functional biological activity of polyethylene wear debris.
Proc Inst Mech Eng H, 215 (2001), pp. 127-132
[27]
R.L. Illgen 2nd, T.M. Forsythe, J.W. Pike, M.P. Laurent, C.R. Blanchard.
Highly crosslinked vs conventional polyethylene particles – an in vitro comparison of biologic activities.
J Artrhroplasty, 23 (2008), pp. 721-731
[28]
A.L. Galvin, J.L. Tipper, L.M. Jennings, M.H. Stone, Z.M. Jin, E. Ingham, et al.
Wear and biological activity of highly crosslinked polyethylene in the hip under low serum protein concentrations.
Proc Inst Mech Eng H, 221 (2007), pp. 1-10
[29]
J.H. Dumbleton, J.A. D’Antonio, M.T. Manley, W.N. Capello, A. Wang.
The basis for a second-generation highly cross-linked UHMWPE.
Clin Orthop Rel Res, 453 (2006), pp. 265-271
[30]
A.L. Galvin, L.M. Jennings, J.L. Tipper, E. Ingham, J. Fisher.
Wear and creep of highly crosslinked polyethylene against cobalt chrome and ceramic femoral heads.
Proc Inst Mech Eng H, 224 (2010), pp. 1175-1183
[31]
A.P. Stoller, T.S. Johnson, O.O. Popoola, S.M. Humphrey, C.R. Blanchard.
Highly crosslinked polyethylene in posterior-stabilized total knee arthroplasty. In vitro performance evaluation of wear, delamination, and tibial post durability.
J Arthroplasty, 26 (2011), pp. 483-491
[32]
Y. Minoda, M. Aihara, A. Sawaka, S. Fukuoka, K. Hayakawa, M. Tomita, et al.
Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty.
[33]
J.T. Hodrick, E.P. Severson, D.S. McAlister, B. Dahl, A.A. Hofmann.
Highly crosslinked polyethylene is safe for use in total knee arthroplasty.
Clin Orthop Relat Res, 466 (2008), pp. 2806-2812
[34]
A.L. TipperJ.L. Galvin, S. Williams, H.M. McEwen, M.H. Stone, E. Ingham, J. Fisher.
Isolation and characterization of UHMWPE wear particles down to ten nanometers in size from in vitro hip and knee joint simulators.
J Biomed Mater Res A, 78 (2006), pp. 473-480
[35]
G.I. Howling, P.I. Barnett, J.L. Tipper, M.H. Stone, J. Fisher, E. Ingham.
Quantitative characterization of polyethylene debris isolated from periprosthetic tissue in early failure knee implants and early and late failure Charnley hip implants.
J Biomed Mater Res, 58 (2001), pp. 415-420
[36]
T.R. Green, J. Fisher, M. Stone, B.M. Wroblewski, E. Ingham.
Polyethylene particles of a critical size are necessary for the induction of cytokines by macrophages in vitro.
Biomaterials, 19 (1998), pp. 2297-2302
[37]
A. Kobayashi, W. Bonfield, Y. Kadoya, T. Yamac, M.A. Freeman, G. Scott, et al.
The size and shape of particular polyethylene wear debris in total joint replacements.
Proc Inst Mech Eng H, 211 (1997), pp. 11-15
[38]
D. MacDonald, A. Sakona, A. Ianuzzi, C.M. Rimnac, S.M. Kurtz.
Do first-generation highly crosslinked polyethylenes oxidize in vivo.
Clin Orthop Relat Res, 469 (2011), pp. 2278-2285

Please cite this article as: Hinarejos P, et al. Partículas de polietileno en líquido sinovial tras artroplastia de rodilla con un polietileno convencional o uno altamente entrecruzado. Estudio preliminar. Rev Esp Cir Ortop Traumatol. 2012;56:210-15.

Copyright © 2011. SECOT
Article options
Tools
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos