Buscar en
Revista Argentina de Microbiología
Toda la web
Inicio Revista Argentina de Microbiología Molecular detection of SARS-CoV-2 in Argentina: Evaluation of alternative diagno...
Información de la revista
Vol. 55. Núm. 3.
Páginas 206-213 (Julio - Septiembre 2023)
Compartir
Compartir
Descargar PDF
Más opciones de artículo
Visitas
759
Vol. 55. Núm. 3.
Páginas 206-213 (Julio - Septiembre 2023)
Original article
Acceso a texto completo
Molecular detection of SARS-CoV-2 in Argentina: Evaluation of alternative diagnostic tools for the decentralization of the diagnosis
Detección molecular de SARS-CoV-2 en Argentina: evaluación de herramientas diagnósticas alternativas para la descentralización del diagnóstico
Visitas
759
Gonzalo Manuel Castroa, Paola Siciliaa, Robertino Gierottoa, Julieta Sosaa, Andrés Marcos Castellarob, María Gabriela Barbásc, María Belén Pisanod,
Autor para correspondencia
mbelenpisano@gmail.com

Corresponding author.
, Viviana E. Réd
a Departamento Laboratorio Central, Ministerio de Salud de la Provincia de Córdoba, Tránsito Cáceres de Allende 421, Córdoba, Argentina
b Centro de Investigaciones en Bioquímica Clínica e Inmunología (CIBICI), CONICET, Facultad de Ciencias Químicas, Universidad Nacional de Córdoba, Av. Medina Allende, X5000 HUA, Córdoba, Argentina
c Secretaría de Prevención y Promoción de la Salud, Ministerio de Salud de la Provincia de Córdoba, Av. Vélez Sarsfield 2311 Ciudad Universitaria, X5016 GCH, Córdoba, Argentina
d Instituto de Virología “Dr. J. M. Vanella”, CONICET, Facultad de Ciencias Médicas, Universidad Nacional de Córdoba, Enfermera Gordillo Gómez s/n, Ciudad Universitaria, CP5016 Córdoba, Argentina
Highlights

  • Tested SARS-CoV-2 molecular assays showed good performance and allowed decentralization.

  • The 9-min RNA extraction was the less time-consuming method yielding quality results.

  • FlashPrep® reagent showed excellent results to perform direct RNA detection.

  • Isothermal amplification assays showed acceptable sensibility and specificity.

  • All the assays were useful for detection of the COVID-19 waves in the central region of Argentina.

Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (5)
Mostrar másMostrar menos
Tablas (1)
Table 1. Sensitivity and specificity of protocols conducted using the isothermal amplification Neokit Plus test or ELA CHEMSTRIP® compared to real time RT-PCR.
Material adicional (2)
Abstract

The rocketing number of COVID-19 cases highlighted the critical role that diagnostic tests play in medical and public health decision-making to contain and mitigate the SARS-CoV-2 pandemic. This study reports the evaluation and implementation of different tests for the molecular detection of SARS-CoV-2 in the central region of Argentina. We evaluated 3 real time RT-PCR kits (GeneFinder COVID-19 Plus RealAmp Kit, DisCoVery SARS-CoV-2 RT-PCR Detection Kit and WGene SARS-CoV-2 RT Detection), 2 nucleic acid extraction methods [MagaBio plus Virus DNA/RNA Purification Kit II (BioFlux), 35-min vs. 9-min], a pre-analytical reagent (FlashPrep®) and 2 isothermal amplification tests (Neokit Plus and ELA CHEMSTRIP®). The order according to the best performance of the 3 real-time RT-PCR kits evaluated was: DisCoVery>GeneFinderTM>WGene. The 2 RNA extraction methods showed similar good results: MagaBio plus Virus RNA Purification Kit II (BioFlux) 9-min was selected due to its faster performance. FlashPrep® reagent showed excellent results to perform direct RNA detection. Isothermal amplification assays showed acceptable sensitivity and specificity values (>80%), except in samples with Ct>30. Our data show optimal real time RT-PCR kits and alternative molecular methods for SARS-CoV-2 diagnostic. These alternative assays proved to be acceptable for their use in adverse contexts, decentralization, and different epidemiological scenarios, for rapid and accurate SARS-CoV-2 detection.

Keywords:
SARS-CoV-2
COVID-19
Diagnosis
Molecular detection
Real time RT-PCR
Isothermal amplification
Resumen

La explosión de casos de COVID-19 resaltó el papel fundamental que desempeñan las pruebas de diagnóstico en la toma de decisiones médicas y de salud pública para contener y mitigar la pandemia de SARS-CoV-2. Este estudio reporta la evaluación y la implementación de diferentes test para la detección molecular de SARS-CoV-2 en la región central de Argentina. Evaluamos tres kits de RT-PCR en tiempo real (GeneFinder COVID-19 Plus RealAmp Kit, DisCoVery SARS-CoV-2 RT-PCR Detection Kit y WGene SARS-CoV-2 RT Detection), dos métodos de extracción de ácidos nucleicos (MagaBio plus Virus DNA/RNA Purification Kit II [BioFlux], 35-min vs. 9-min), un reactivo pre-analítico (FlashPrep®) y dos test de amplificación isotérmica (Neokit Plus and ELA CHEMSTRIP®). El orden de rendimiento de los tres kits de RT-PCR en tiempo real evaluados fue el siguiente: DisCoVery>GeneFinder™>WGene. Los dos métodos de extracción de RNA mostraron buenos y similares resultados; se seleccionó MagaBio plus Virus RNA Purification Kit II (BioFlux) 9-min debido a su rápido tiempo de procesamiento. El reactivo FlashPrep® mostró excelentes resultados para realizar detección directa de RNA. Los ensayos de amplificación isotérmica mostraron valores de sensibilidad y de especificidad aceptables (>80%), excepto en muestras con Ct>30. Nuestros resultados muestran kits de RT-PCR en tiempo real óptimos, como así también métodos moleculares alternativos para el diagnóstico de SARS-CoV-2 que resultan aceptables para su uso en contextos adversos, de descentralización y en diferentes escenarios epidemiológicos, para la detección rápida y precisa del SARS-CoV-2.

Palabras clave:
SARS-CoV-2
COVID-19
Diagnóstico
Detección molecular
RT-PCR en tiempo real
Amplificación isotérmica
Texto completo
Introduction

Since SARS-CoV-2 infection was first detected in Argentina in March 2020, the virus dissemination was exponential, producing a large number of cases8. The rocketing number of COVID-19 cases, which led to the occurrence of three epidemic waves (up to date), highlighted the critical role that diagnostic tests play in medical and public health decision-making to contain and mitigate the SARS-CoV-2 pandemic8. Because Argentina is a federal country, the decisions to fight the pandemic were made by each of the provinces in concordance with the national measures.

In Argentina, as in many other countries, the National Administration of Medicines, Food and Medical Technology granted emergency authorizations for diagnostic tests as they became available and the reference laboratories successively evaluated and implemented them as far as possible, according to accessibility, costs, utility and performance9. To deal with supply chain difficulties that caused shortage of reagents and equipment, clinical laboratories implemented multiple assay tests for SARS-CoV-2 detection in accordance with the availability of each region.

It is important to update and clearly understand the analytical parameters of the available options for SARS-CoV-2 detection, to guide the selection of tests based on certain considerations of the changing epidemiological scenario, such as expanded testing and greater access to them, introduction of vaccines, circulation of new variants, and entry of new tests, among others. With regard to the molecular methods, the licensed platforms use a variety of different primer and probe sets, diverse amplification gene targets, operational and runtime differences, equipment suggested by the manufacturer that is not always available, in addition to the variety of biological samples and extraction methods, which result in differences in analytical sensitivities that must be evaluated.

The objective of this study was to report the experience in the evaluation and implementation of different tests for SARS-CoV-2 molecular detection in the central region of Argentina.

Materials and methods

During this study, different methods for SARS-CoV-2 molecular detection were evaluated according to the availability of kits and samples in our region at different times. The molecular assays evaluated and comparisons performed during this study are shown in Figure 1.

Figure 1.

Molecular assays for SARS-CoV-2 detection tested and compared during this study.

(0,32MB).
Real time PCR

A review of the technical and operational characteristics of the following 10 RT-real time PCR kits for SARS-CoV-2 detection that were available in Argentina during the COVID-19 pandemic was carried out: “in-house” coronavirus RT-real time PCR using probes and primers recommended by the CDC, LightMix Modular SARS-CoV-2 (TIB Molbiol-Syntheselabor GmbH), 2019-nCoV GENESIG (Primerdesign LTD, United Kingdom), TaqMan 2019-nCoV Assay Kit v1 (Applied Biosystems, United States), GeneFinder COVID-19 Plus RealAmp Kit (Osang, Healthcare CO LTD, Korea), VIASURE SARS-CoV-2 Real Time PCR Detection Kit (Certest Biotec, Spain), Novel Coronavirus (2019-nCoV) Detection Kit v2 (Anatolia Diagnostics and Biotechnology Products Inc.), Real-Time Fluorescent RT-PCR Kit for Detecting SARS-2019-nCoV (BGI Genomics Co Ltd, China), DisCoVery SARS-CoV-2 RT-PCR Detection Kit (Safecare Biotech Hangzhou Co Ltd, China) and WGene SARS-CoV-2 RT Detection Kit (WIENER SAIC, Argentina).

We analyzed and compared the following parameters: target genes, internal control (CI), single reaction/multiplex, sample volume, reagent yield, required equipment and reaction time.

For the comparison of GeneFinder COVID-19 Plus RealAmp Kit vs. DisCoVery SARS-CoV-2 RT-PCR Detection Kit, 64 RNA samples, previously determined for SARS-CoV-2 detection, were analyzed (positive samples n=43; negative samples n=21).

For the comparison of DisCoVery SARS-CoV-2 RT-PCR Detection Kit vs. WGene SARS-CoV-2 RT Detection Kit, 94 RNA samples, previously determined for SARS-CoV-2 detection, were analyzed [positive samples n=76 with Cts for the N-gene during initial diagnosis: <20 (n=19), 20–30 (n=29), >30 (n=28); negative samples n=18].

Nucleic acid extraction test

For RNA extraction, 2 automatized methods were tested: (1) MagaBio plus Virus DNA/RNA Purification Kit II (BioFlux), 35min and (2) MagaBio plus Virus RNA Purification Kit II (BioFlux), 9min, using GenePure Pro Nucleic Acid Purification System NPA-32P (Bioer). A panel of 64 positive SARS-CoV-2 oropharingeal swab samples (previously determined by RT-real time PCR DisCoVery SARS-CoV-2 RT-PCR Detection Kit) were simultaneously tested by these methods. Briefly, 300μl of each sample were added to a well in a 96 well plate (one sample per well) and introduced into the equipment where the automated extraction was performed. For the MagaBio plus Virus DNA/RNA Purification Kit II, 10μl of proteinase K were added to each well containing the sample.

FlashPrep® treatment followed by RT-real time PCR

A panel of 74 oropharyngeal swab samples obtained in physiological saline solution, previously tested for SARS-CoV-2 detection by RT-real time PCR using DisCoVery SARS-CoV-2 RT-PCR Detection Kit [positive samples n=58, with Cts for the N-gene during initial diagnosis: <20 (n=4), 20–30 (n=42), >30 (n=12); negative samples n=16], was utilized. Briefly, 90μl of each sample were incubated with 10μl of the FlashPrep® RNA SARS-CoV-2 reagent (Inbio Highway, Argentina) for 15min at 55°C and 5min at 98°C for viral inactivation and rapid release of SARS-CoV-2 RNA, according to the manufacturer's specifications. Then, 8μl (suggested by the manufacturer) and 5μl (modified according to the requirements of the real time RT-PCR kit) of the obtained RNA were used in parallel for the RT real time-PCR.

Isothermal amplification using Neokit Plus

A hundred and four (104) randomly selected saliva samples were processed for: 1-RNA automatized extraction with MagaBio plus Virus RNA Purification Kit II (BioFlux) 9min, to be subjected to molecular SARS-CoV-2 detection by RT-real time PCR (using DisCoVery SARS-CoV-2 RT-PCR Detection Kit) and isothermal amplification using COVID19 Neokit Plus (Neokit S.A.S, Argenitna) (revealed with a colorimetric reaction); and 2-direct molecular detection by isothermal SARS-CoV-2 RNA amplification with COVID19 Neokit Plus (Neokit S.A.S, Argentina).

Briefly, in case No. 1 (starting from RNA), 5μl of RNA of each sample were added to 35μl of the mix provided in the kit, containing primers and enzymes, and incubated for 60min at 64°C. In case No. 2 (starting from saliva samples, without the extraction process), 10μl of each sample were added to 40μl of a lysis buffer and incubated for 8min at 97°C. After that, 10μl of this lysate were added to another tube with 30μl of the reaction buffer and incubated for 60min at 64°C. After that time, in both cases, No. 1 and No. 2, the colour of the samples was observed, and absorbance was measured using the provided equipment. Blue samples were considered positive, while violet samples were negative. In addition, graphs of the increase in absorbance were obtained.

A reagent control, as well as positive and negative controls were included in each run.

Isothermal amplification using ELA CHEMSTRIP®

A total of 91 RNA samples (extracted using MagaBio plus Virus RNA Purification Kit II 9min, as previously described), obtained from oropharyngeal swabs and previously tested for SARS-CoV-2 detection by RT-real time PCR using DisCoVery SARS-CoV-2 RT-PCR Detection Kit [positive samples n=60, with Cts for the N-gene during initial diagnosis: <20 (n=23), 20–30 (n=22), >30 (n=23); negative samples n=23], were analyzed for isothermal amplification using an Easy Loop Amplification (ELA) and the Gen cap kit (ELA CHEMSTRIP® COVID-19, CHEMTEST ARGENTINA S.A., Argentina), following the manufacturer's instructions. Briefly, 5μl of RNA were added to a mix containing Mix ELA COVID19, primers and enzyme (reagents provided by the manufacturer), and incubated for 60min at 60°C. Then, 300μl of sample diluent were added, and the mixture was put onto the strip to perform immunochromatography. The result was read after 10min. Positive and negative controls were included. An internal process control (to be added in the first step) is included in the kit.

Statistical analyses

Statistical analyses were carried out using Epidat 3.1 program3. Sensitivity and specificity were calculated in most of the cases and expressed as percentages. Kappa coefficient (κ) was calculated to determine how similarly the tests performed on a 1–1 basis. A κ of 0.8 or greater was considered a strong agreement6.

ResultsReal time PCR

After the technical and operational revision, GeneFinder, DisCoVery and WGene were selected for further comparisons as they exhibited the most suitable characteristics for the laboratory requirements: presence of internal control, multiplex reaction, detection of more than one gene, good performance and reduced processing time (Supplementary Table 1).

A 100% concordance was observed when comparing GeneFinder vs. DisCoVery, as well as DisCoVery vs. WGene (κ=1 in both cases), with an average ΔCt value of 0.8 (range=0–2.0) and 3.3 (range=1.2–5.0), respectively. In the first case, Ct values were very similar (almost the same as for the N gene) (Fig. 2), and in only 2 samples the difference in Cts for the N gene was as high as 2 (samples No. 4 and 12, Fig. 2A). Cts obtained for WGene were between 1.2 and 5.0 higher than those obtained by DisCoVery, mainly for the N gene (Fig. 3). Due to the aforementioned (lower Ct values), added to the fact that Discovery is less time-consuming, it was selected as the reference test for the following evaluations.

Figure 2.

Ct values obtained when processing SARS-CoV-2 RNA samples using the GeneFinder (blue) and DisCoVery (red) kits for the N gene (A) and the IC (B).

(0,41MB).
Figure 3.

Ct values obtained when processing SARS-CoV-2 RNA samples using the DisCoVery (blue) and WGene (red) kits for the N gene (A) and the IC (B).

(0,47MB).
Nucleic acid extraction tests

Results of the comparison of the 2 RNA extraction protocols showed 100% concordance for the 2 SARS-CoV-2 genes tested (N and ORF1ab) as well as the IC (all the positive samples obtained with one method were also positive with the other method) (κ=1) (Fig. 4). For both the N and the ORF1ab genes, only one sample showed a ΔCt as high as 5.5 (gene N) and 6.6 (ORF1ab gene) (sample No. 7, Fig. 4), with a better performance for the 35-min kit. For the rest of the samples, Cts obtained with the 35-min test for these genes were slightly lower than those obtained with the 9-min test (Fig. 4). The overall mean difference in Ct values was: for the N gene=1.3 (range 0–5.5), for the ORF1ab gene=1.5 (range 0–6.6).

Figure 4.

Ct values obtained for the N (A) and ORF1ab (B) genes, and the IC (C), when processing oropharyngeal swab samples for SARS-CoV-2 detection using the 2 RNA extraction protocols of 35min (blue) and 9min (red).

(0,58MB).
FlashPrep® treatment followed by RT-real time PCR

Figure 5 shows the correlation of SARS-CoV-2 molecular detection using FlashPrep® RNA SARS-CoV-2 reagent in the 2 volumes tested (5μl and 8μl) vs. classical RNA extraction (using the 9-min protocol) in the 74 samples analyzed. Excellent correlation was observed between the results, without substantial variation of the Cts obtained using both methods, both starting from 5μl of RNA (ΔCt=0.2; range=0.1–5.7) as well as starting from 8μl of RNA (ΔCt=0.7, range=0–6.1).

Figure 5.

Ct values obtained for the N (A) and ORF1ab (B) genes, and the IC (C), when processing oropharyngeal swab samples for SARS-CoV-2 detection using FlashPrep® RNA SARS-CoV-2 reagent in the 2 volumes tested [5μl (blue) and 8μl (red)] vs. classical RNA extraction (performed with the 9-min protocol) (green).

(0,61MB).

When comparing the results obtained using the 2 starting RNA volumes (5μl vs. 8μl), a high similarity was observed in the Ct values, with a very low range between them (range=0–1.5).

Isothermal amplification using Neokit Plus

The Neokit Plus isothermal amplification assay was evaluated in saliva samples, following 2 protocols: 1 – molecular detection of RNA extracted from the samples (using the 9-min protocol) and 2 – molecular detection directly from the samples (direct amplification). Both protocols were compared to RNA detection by RT real time PCR (from RNA extracted from the same samples). The sensitivity and specificity values obtained for both evaluated methods were higher than 80%, which was considered acceptable (Table 1A and Supplementary Table 2), although the Neokit Plus assay applied to SARS-CoV-2 molecular detection from RNA exhibited higher performance.

Table 1.

Sensitivity and specificity of protocols conducted using the isothermal amplification Neokit Plus test or ELA CHEMSTRIP® compared to real time RT-PCR.

  A – Neokit PlusB – ELA CHEMSTRIP®
  Direct detection(saliva samples)  RNA extracted from saliva  ELA CHEMSTRIP®  ELA CHEMSTRIP® in samples with Cts>30 
Sensitivity  82.4% (68.1–96.6%)  94.1% (84.7–100.0%)  88.2% (79.8–96.6%)  65.2% (43.6–86.9%) 
Specificity  80.0% (69.9–90.1%)  87.1% (78.6–95.7%)  100.0% (97.8–100.0%)  100.0% (97.8–100.0%) 

Values are expressed as percentages with a 95% confidence interval (CI).

ELA CHEMSTRIP® isothermal amplification

RNA samples extracted from oropharyngeal swabs using the 9-min protocol with different Ct values, were analyzed for SARS-CoV-2 detection by ELA CHEMSTRIP® isothermal amplification and compared to results obtained by RT-real time PCR in the same samples. Sensitivity and specificity presented acceptable values (higher than 80%) (Table 1B and Supplementary Table 3A).

In all samples with Cts<30, 100% concordance was observed in the results obtained by ELA CHEMSTRIP® and real time PCR. However, in samples with Cts>30, the sensitivity of the tested assay decreased considerably (Table 1B and Supplementary Table 3B).

Discussion

Some of the greatest problems of the pandemic are related to the lack of knowledge about the virus, the uncertainty and lack of resources, which pose new challenges and lead us to a constant generation of information in the field of health, including diagnosis. As the pandemic progressed, the number of COVID-19 cases increased with the identification of new variants, and a number of obstacles have arisen, such as the shortage of supplies (swabs, disposable material, PCR reagents, RNA isolation kits, etc.), the need for specific equipment, qualified human resources, increased testing to control and isolate cases, and the introduction of new methodologies that must be validated for all the virus variants9. This scenario raises the need to be prepared for all situations, with useful diagnostic options and strategies that contribute to the control of the outbreaks. In this context, in the reality of a developing country (such as Argentina), this work describes the experience acquired in the evaluation and implementation of different strategies and methods for the detection of SARS-CoV-2.

As in the rest of the world, SARS-CoV-2 began to be tested using in-house PCRs, with primers designed by world reference centers2. Later, commercial PCR tests became available for the amplification of different SARS-CoV-2 genomic targets, as well as isothermal tests and also new formats and reagents for RNA extraction1.

At the beginning of the pandemic, only 10 kits for SARS-CoV-2 detection by real time RT-PCR were available in Argentina1. Of these, GeneFinder proved to be the most reliable one, and of choice for the requirements of a laboratory that processed between 3000 and 5000 samples per day, such as ours. The selection was performed based on some characteristics: it had internal control, amplified more than one gene, was a multiplex reaction (all the reactions were carried out simultaneously), and achieved very good performance. After some time, the Discovery kit was introduced in Argentina and started to be commercialized. It exhibited similar general characteristics to GeneFinder and a high degree of concordance, but with a shorter reaction time, which is why it was selected as the gold standard for molecular diagnosis and was used for the evaluation of the rest of the methods presented here. Although the WGene kit did not show to have superior characteristics to Discovery, its performance was acceptable, therefore, it could be used in case of reagent shortage. During 2020, 2 isothermal amplification-based kits were developed and manufactured in Argentina (Neokit and ELA CHEMSTRIP®), which were tested during this study, as alternative diagnostic methodologies. They exhibited overall acceptable values for sensitivity and specificity (higher than 80%), proposing them as alternative diagnostic tests in places where specific infrastructure, equipment and/or trained personnel are not available. In addition, they have the advantage of low cost and short processing time (min), which allows to carry out tests on a larger scale10. However, the performance of the assays decreased when they were performed from the original sample (without RNA extraction in the case of Neokit) or for samples with low viral load (with Cts>30), and ELA CHEMSTRIP® result was operator-dependent (since the result must be read by interpreting the presence/absence of bands in a strip), which could lead to false results. Therefore, it is always preferable to use real time RT-PCR as a first-choice technique, without ruling out the use of isothermal kits in contexts that present the aforementioned limitations.

For SARS-CoV-2 detection, nasopharyngeal (NP) specimens are the preferred choice samples for swab-based testing, followed by oropharyngeal (OP) specimens9; however, they must be collected by a trained healthcare worker, have an increased cross-infection risk, possibly causing discomfort, coughing and even bleeding in patients, not being so desirable for serial viral monitoring5. As an alternative option, salivary tests have demonstrated similar performance to NP and/or OP swabs for SARS-CoV-2 diagnosis4. They are a simple and non-invasive alternative, samples can be self-collected outside hospitals, without the need of a health-care professional, reducing nosocomial transmission risk, test waiting time, and transport and storage costs5. During this study saliva samples were tested by real time RT-PCR and isothermal amplification (direct detection and detection of RNA extracted from these samples), with acceptable performance (if the requirements for sample collection are met, at least 1 hour fast, no use of mouthwash or tooth brushing, among others), showing that saliva can be used for diagnosis, mainly in cases in which there is no personnel to take samples when there is a shortage of supplies (swabs), or in certain populations in which sampling is difficult, such as children11.

Better results were obtained using extracted RNA in the SARS-CoV-2 detection by isothermal amplification using Neokit Plus (lower performance was registered when starting from the original sample). On the other hand, when using the FlashPrep® RNA SARS-CoV-2 reagent, excellent results were observed for direct detection (from the original sample), even when using 5μl as starting volume for the real time RT-PCR. The storage conditions of the samples are considered a determining factor that may have influenced the performance of these assays. RNA extraction has been reported to be successfully bypassed when samples are stored in a universal transport medium (UTM) or in molecular water but not when samples are stored in saline solution and in Hanks medium7. On the contrary, the FlashPrep® RNA SARS-CoV-2 reagent is prepared for the processing of samples in physiological solution and does not allow the use of commercial tubes with viral transport medium. This is a factor to be considered when using this type of direct detection assays (without the extraction process), added to the need to have a thermoblock or thermal cycler for the incubation step. Despite these limitations, these techniques are useful in certain contexts, with scarce resources, lack of personnel and equipment to carry out the extraction.

Two protocols for RNA extraction with different processing times (35 and 9min) were tested, and a high concordance in the results was observed. Although the 9-min protocol presented some differences in the ΔCt that could impact those cases in which the Cts are high, this difference would be negligible when evaluating other parameters that would justify its use, such as its speed, usefulness in cases of massive testing, and centralized diagnostic centres, since it allows to extract a greater number of samples in a shorter period of time, increasing the useful life of the equipment and reducing working hours Therefore, this method proved to be an excellent choice for SARS-CoV-2 RNA extraction.

Conclusion

In conclusion, we presented an exhaustive analysis of methods for molecular detection of SARS-CoV-2, showing a great variety of available assays, which can be used in different contexts and diverse epidemiological scenarios. Although the reference technique is real time RT-PCR, it requires skilled personnel and a specialized thermal cycler, taking several hours to obtain results. Our results show other molecular methods or options, such as isothermal amplification or reagents to avoid RNA extraction, which have shown to be acceptable for its use in adverse contexts for rapid and accurate SARS-CoV-2 detection.

Funding

This research has not received any specific grant from agencies in the public, commercial, or non-profit sectors.

Conflict of interest

The authors declare that they have no conflicts of interest.

Acknowledgements

We thank the staff of the Laboratorio Central de la Provincia de Córdoba, for helping obtaining the samples.

Appendix A
Supplementary data

The following are the supplementary data to this article:

References
[2]
Centre for Disease Control and Prevention (CDC). https://www.cdc.gov/spanish/index.html [accessed 17.01.22].
[3]
Epidat 3.1 – Xunta de Galicia – Consellería de Sanidade y el Servicio Gallego de Salud.
(2006),
[4]
T. Fukumoto, S. Iwasaki, S. Fujisawa, K. Hayasaka, K. Sato, S. Oguri, K. Taki, S. Nakakubo, K. Kamada, Y. Yamashita, S. Konno, M. Nishida, J. Sugita, T. Teshima.
Efficacy of a novel SARS-CoV-2 detection kit without RNA extraction and purification.
Int J Infect Dis, 98 (2020), pp. 16-17
[5]
R.C. Madeiros da Silva, L.C. Nogueira Marinho, D.N. de Araújo Silva, K. Costa de Lima, F.Q. Pirih, A.R. Luz de Aquino Martins.
Saliva as a possible tool for the SARS-CoV-2 detection: a review.
Travel Med Infect Dis, 38 (2020), pp. 101920
[6]
McHugh.
Interrater reliability: the kappa statistic.
Biochem Medica, 22 (2012), pp. 276-282
[7]
N. Merindol, G. Pépin, C. Marchand, M. Rheault, C. Peterson, A. Poirier, C. Houle, H. Germain, A. Danylo.
SARS-CoV-2 detection by direct rRT-PCR without RNA extraction.
J Clin Virol, 128 (2020), pp. 104423
[8]
National Ministry of Health of Argentina. Epidemiological reports. https://www.argentina.gob.ar/coronavirus/informes-diarios/reportes [accessed 17.01.22].
[9]
World Health Organization (WHO). https://www.who.int/es [accessed 17.01.22].
[10]
C. Yan, J. Cui, L. Huang, B. Du, L. Chen, G. Xue, S. Li, W. Zhang, L. Zhao, Y. Sun, H. Yao, N. Li, H. Zhao, Y. Feng, S. Liu, Q. Zhang, D. Liu, J. Yuan.
Rapid and visual detection of 2019 novel coronavirus (SARS-CoV-2) by a reverse transcription loop-mediated isothermal amplification assay.
Clin Microbiol Infect, 26 (2020), pp. 773-779
[11]
R. Yee, T.T. Truong, P.S. Pannaraj, N. Eubanks, E. Gai, J. Jumarang, L. Turner, A. Peralta, Y. Lee, J. Dien Bard.
Saliva is a promising alternative specimen for the detection of SARS-CoV-2 in children and adults.
J Clin Microbiol, 59 (2021), pp. e02686-e2720
Copyright © 2023. Asociación Argentina de Microbiología
Opciones de artículo
Herramientas
Material suplementario
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos