Hand dynamometry: Does the device used for measurement matter?

Dinamometría de la mano: ¿importa el dispositivo utilizado en la medición?
Visits
2283
Diana García Sastre, Begoña Pintor de la Maza, María López Melgar, María D. Ballesteros Pomar
Corresponding author
Endocrinology and Nutrition, Complejo Asistencial Universitario de León, León, Spain
This item has received
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (3)
fig0005
fig0010
fig0015
Tables (2)
Table 1. Intraclass correlation of the 4 dynamometers.
Tables
Table 2. Bias and limits of agreement of the 3 dynamometers vs the gold standard.
Tables
Abstract
Introduction

Handgrip dynamometers measure muscle strength through the grip strength of the dominant hand. The purpose of this study is to compare the Dynx, Saehan, and Smedlay dynamometers with the Jamar dynamometer (gold standard) by evaluating the correlation and concordance of their measurements to determine if it is indifferent to use different dynamometers.

Material and methods

Cross-sectional study conducted between 2020 and 2021.

We evaluated a group of patients (GP) from the Endocrinology and Nutrition consultations and a control group (CG) of health personnel.

Dominant hand grip strength was measured with four different dynamometers Jamar, Dynx, Saehan and Smedlay, in random order. Statistical analysis was performed with SPSS using Student's t-test, intraclass correlation coefficient (ICC), and Bland-Altman plots.

Results

A total of 124 subjects were included; 91 GP (60.4% women), mean age 54.9 (SD14.2) years. ICC showed the best correlation for Saehan in GP (0.977 p < 0.001) and in GC (0.960 p < 0.001). Dynx correlated better in GP than in GC (0.967 vs. 0.900, p < 0.001) and Smedlay had the worst correlation in GC (0.812 p < 0.001). The dynamometer with the lowest bias (-0.39, SD 3.11) and better limits of agreement (LOA) in both groups (GP -0.39, GC -0.51) was the Saehan

Conclusion

Despite good measurement correlation, there is bias and insufficient agreement between different dynamometers, so they should not be used interchangeably.

Keywords:
Dynamometer
Hand Grip Strength (HGS)
Sarcopenia
Intraclass correlation
Bland Altman
Resumen
Introducción

Los dinamómetros de empuñadura miden la fuerza muscular a través de la fuerza de agarre de la mano dominante. El objetivo de este estudio es comparar los dinamómetros Dynx, Saehan y Smedlay con el dinamómetro Jamar (gold estándar) evaluando la correlación y concordancia de sus medidas, comprobando si es indiferente emplear distintos dinamómetros

Material y métodos

Estudio transversal realizado entre 2020 y 2021.

Se evaluó a un grupo de pacientes (GP) procedentes de las consultas de Endocrinología y Nutrición y a un grupo control (GC) de personal sanitario.

Se midió la fuerza de agarre de la mano dominante con cuatro dinamómetros diferentes Jamar, Dynx, Saehan y Smedlay, en orden aleatorio. El análisis estadístico se realizó con SPSS, utilizando la prueba T de Student, el coeficiente de correlación intraclase (ICC), y gráficos Bland-Altman.

Resultados

Se incluyeron 124 personas; 91 GP (60,4% mujeres), edad media de 54,9 (DE14,2) años. El ICC mostró la mejor correlación para Saehan en el GP (0.977 p < 0.001) y en el GC (0.960 p < 0.001). Dynx correlacionó mejor en el GP que en GC (0.967 vs 0.900, p < 0.001) y Smedlay obtuvo la peor correlación en el GC (0.812 p < 0.001). El dinamómetro que menor sesgo obtuvo (-0.39, SD 3.11) y mejores medias en los límites de acuerdo (LOA) en ambos grupos (GP -0.39, GC -0.51) fue Saehan.

Conclusión

A pesar de una buena correlación de la medida, existe un sesgo y una concordancia inadecuada entre distintos dinamómetros, por lo que no deben utilizarse indistintamente.

Palabras clave:
Dinamómetro
Fuerza de agarre (HGS)
Sarcopenia
Correlación intraclase
Bland Altman
Full Text
Introduction

Handheld dynamometers are portable devices, simple to use without requiring specific training for the professional performing the test, that allow for the evaluation of muscle strength through the grip strength of the dominant hand (hand grip strength or HGS).1

Low grip strength predicts unfavorable situations in many patients, such as a longer length of stay, functional problems, lower quality of life, and increased mortality, as it is used to assess maximum muscle strength.2 Low levels of maximum muscle strength can indicate the presence of sarcopenia, which influences the evolution of these patients.3

Due to its ease of use, grip strength is routinely used in standard clinical practice. It is important to use a calibrated handheld dynamometer in good condition to obtain interpretable data and compare them with the reference population.

Currently, different models of handheld dynamometers can be found. They can be electric or analog, hydraulic or mechanical, and have different grips. The American Society of Hand Therapists endorses the Jamar® Hydraulic dynamometer as the reference device.4

Several studies have documented the differences between the Jamar dynamometer and other dynamometer models, both analog and electric. To study the reliability and validity of other devices on the market, Orit Shechtman et al. compared the Jamar dynamometer with the Dynx in healthy subjects (20–40 years old) and found significant differences between the grip scores obtained (± 6.222 kg; p = 0.014).5 After examining these results, our study aims to expand the range of compared devices in the current literature. Handheld dynamometers measure muscle strength through the grip strength of the dominant hand. We observed in routine clinical practice the differences in measurements depending on the dynamometer used and thus defined the objective of this study: to compare the Dynx, Saehan, and Smedlay dynamometers with the Jamar dynamometer (reference device) to evaluate the correlation and agreement of their measurements and to check if it is indifferent to use different dynamometers.

Material and methods

We conducted a cross-sectional study was conducted from October 2020 through April 2021. Patients included in the study were referred from the Nutrition Unit (outpatient follow-up) of the Endocrinology and Nutrition Department of Hospital Universitario de León (León, Spain). They were offered to participate in the study and were classified as the patient group (PG). The control group (CG) included health care professionals younger than 40 years.

The sample size calculation was performed using the Ene 3.0 software, Glaxo-Smith-Kline, Biometrics Department®, 1992–2017, based on the study by Guerra et al.6 To achieve a statistical power of 80.0% and detect differences in the null hypothesis test H₀: m = 17.60, we used the bilateral Student's t-test for one sample, considering a significance level of 5.00%, assuming the mean is 14.6 units and the population standard deviation is 7.8 units. Based on all of this, it was concluded that the inclusion of 55 experimental units in the study was necessary. Patients and the control group gave their prior written informed consent form following the considerations outlined in the Declaration of Helsinki. We emphasized respect for participants and the protection of their rights and well-being, ensured data confidentiality, minimized risks, and guaranteed that potential benefits outweigh potential harms (Complejo Asistencial de León Research Ethics Committee, registration No. 2138).

General data were collected to characterize both groups, including sex, age, height, weight, body mass index (BMI), and diagnosis in the PG. HGS was measured with 4 different dynamometers: 2 electric dynamometers (Dynx, Akern® and Saehan®) and 2 analog dynamometers (Jamar® Hydraulic, the reference, and Smedlay®).

Measurements were taken on the dominant hand, and the best of 3 consecutive measurements was recorded. Patients were seated with shoulders at 0 ° abduction and neutral rotation, elbows at 90 ° flexion unsupported, and forearms in neutral pronation/supination. The order in which the dynamometer was given to the subjects was randomized each time using a website (https://pinetools.com/es/aleatorizar−lista).

Statistical analysis was performed using SPSS® software (version 15.0, Chicago, IL, USA). Continuous data are expressed as mean (standard deviation, SD). A comparison of z-scores relative to the mean was performed to discern how far data points are from the mean in relative terms. Student's t-test was used to evaluate differences within groups. A p-value <0.05 was considered significant. Reliability analysis was performed using the intraclass correlation coefficient (ICC), considering low agreement if ICC was <0.40, regular/good if 0.41−0.75, and very good if >0.75. Bland-Altman plots were also created to compare 2 measurement techniques for the same quantitative variable and estimate the precision of the results, as it serves to represent the difference between the results of measurements taken by different measurement methods versus the mean of both, and to identify any systematic bias or random error in the data.

Results

A total of 124 individuals were included in the study; 33 subjects were included in the CG (78.8% women) with a mean age of 32.3 years (SD, 7.3). In the PG, 91 patients were included (60.4% women) with a mean age of 54.9 years (SD, 14.2).

In the PG, mean height was 1.65 m (SD, 0.11), weight was 79.8 kg (SD, 25.9), and BMI was 29.05 kg/m² (SD, 8.37). In the CG, mean height was 1.69 m (SD, 0.08), weight was 63.6 kg (SD, 11.4), and BMI was 22.30 kg/m² (SD, 3.41). The PG had the following diagnoses: previous bariatric surgery (28.6%), obesity (24.2%), cancer (15.4%), digestive disease (11.0%), others (20.8%).

HGS results in the PG were 25.73 kg (SD, 9.71) with Dynx, 28.31 kg (SD, 10.04) with Saehan, 27.91 kg (SD, 10.58) with Jamar, and 29.59 kg (SD, 10.02) with Smedlay. HGS results for the CG were 30.10 kg (SD, 9.20) with Dynx, 31.98 kg (SD, 8.56) with Saehan, 31.47 kg (SD, 8.61) with Jamar, and 32.84 kg (SD, 7.80) with Smedlay. ICC results are shown in Table 1.

Table 1.

Intraclass correlation of the 4 dynamometers.

Dynamometer (kg)  CG  PG 
Jamar®  Reference device  Reference device 
Saehan®  0.960 (0.919–0.980); p < 0.001  0.977 (0.965–0.985); p < 0.001 
Dynx, Akern®  0.900 (0.798–0.951); p < 0.001  0.967 (0.951–0.979); p < 0.001 
Smedlay®  0.812 (0.620–0.907); p < 0.001  0.963 (0.944–0.976); p < 0.001 

CG: control group; PG: patient group.

The complete results of the mean difference of the z-scores are shown in Table 2.

Table 2.

Bias and limits of agreement of the 3 dynamometers vs the gold standard.

  PG Jamar vs other dynamometers  PG LOA  CG Jamar vs other dynamometers  PG LOA 
Jamar®  Reference device  Reference device  Reference device  Reference device 
Saehan®  −0.39 (SD, 3.11)  Upper: 5.70 Lower: −6.47  −0.51 (SD, 3.37)  Upper: 6.11 Lower: −7.13 
Dynx, Akern®  2.19 (SD, 3.61); p < 0.001  Upper: 5.95 Lower: −9.31  1.37 (SD, 5.37)  Upper: 11.44 Lower: −14.16 
Smedlay®  −1.68 (SD, 3.89); p < 0.001  Upper: 9.26 Lower: −4.88  −1.37 (SD, 6.53)  Upper: 11.90 Lower: −9.16 

CG: control group; PG: patient group; LOA: limits of agreement; SD: standard deviation.

To estimate the agreement between the Jamar® dynamometer and the other 3 dynamometers, a Bland-Altman plot was prepared. The results of each comparison are shown in Figs. 1–3.

Figure 1.

Bland-Altman plot of Jamar vs Dynx.

CG: control group; PG: patient group; LOA: limits of agreement; Lower: inferior; Mean: media; Upper: superior.

Figure 2.

Bland-Altman plot of Jamar vs Smedlay.

CG: control group; PG: patient group; LOA: limits of agreement; Lower: inferior; Mean: media; Upper: superior.

Figure 3.

Bland-Altman plot of Jamar vs Saehan.

CG: control group; PG: patient group; LOA: limits of agreement; Lower: inferior; Mean: media; Upper: superior.

Discussion

The purpose of this study is to compare several dynamometers with the Jamar, considered the gold standard, attempting to define standards for performing muscle strength tests through dominant hand grip strength. An excellent correlation is found, although with significant biases and dispersion, especially in the patient group. The results indicate that dynamometers cannot be used interchangeably due to a lack of adequate agreement.

The expert consensus on sarcopenia3 recommends measuring handgrip strength with a single dynamometer, but different devices are available on the market, and few studies in the literature compare more than 2 or 3 dynamometers regarding the existing measurement differences between them. Therefore, comparing them with the reference standard seems relevant when validating results obtained in other studies with dynamometers other than the Jamar.

Although specific cut-off points have been used for diagnosis, the use of different dynamometers can affect the measurement and, directly, the diagnosis. This study showed that the electric Dynx dynamometer and the analog Smedlay dynamometer have results that correlate well with the reference (Jamar) but with a clinically relevant bias and wide margins of disagreement. The Saehan device, with the same mechanism as Jamar, differing only in providing digital instead of analog measurements, shows little difference and good correlation with Jamar. The other 2 dynamometers present clinically relevant differences. The intraclass correlation among all devices is very good, but Dynx overestimates the measurement by more than 2 kg in the patient group, and Smedlay underestimates it by almost 2 kg, which should be considered when evaluating cut-off points and diagnoses of low muscle strength. In the case of the CG, the difference is smaller, between +1.37 for Dynx and −1.37 for Smedlay.

In 2019,7 Ha et al. conducted a systematic review of 34 epidemiological studies to evaluate the number of times HGS should be repeated, as it is still a controversial topic. Most of the included studies performed 2 or 3 measurements, according to the recommendations of the American Society of Hand Therapists, a principle that we maintained in our methodology.

In 2022,8 Huang et al. compared 2 dynamometers, the Camry, with a grip very similar to the Smedlay, and the Jamar. They measured HGS in a total of 1,064 healthy adults aged between the ages of 50 and 90 years (65.1% women), randomizing the test order with the 2 dynamometers, with a 10-minute interval between each test. ICC values between the 2 devices and Bland-Altman analysis were calculated to evaluate reliability and validity between both groups. Mean HGS values were 25.0 ± 7.9 kg with Jamar and 24.6 ± 7.5 kg with Camry. ICC values between the devices were 0.815 to 0.854. Systematic bias was expressed as the mean difference between 2 methods using Bland-Altman plots. The 95% limits of agreement (LOA) were defined as bias ± 1.96 SD, (p < 0.05) and indicated statistical significance. HGS results showed a high association between both Jamar and Camry dynamometers.

Smedlay and Jamar were also compared by Benton et al. in 2022,9 who analyzed measurements from 67 older patients (76.2 ± 0.9 years). They observed strong intraclass correlations (p < 0.001) for the 2 devices (Jamar = 0.98; Smedley = 0.96). HGS results collected by Jamar were superior to those of Smedlay (27.4 ± 1.4; 23.4 ± 1.1 kg, respectively), which does not correlate with the results obtained in this study, as Smedlay was superior to Jamar in both the CG and the PG. As in our study, the Bland-Altman diagram between Smedlay and Jamar showed poor agreement between both dynamometers, despite a good intraclass correlation of 0.96.

In a recent study from December 2023, Carli Lessof et al.10 conducted a comparative analysis of HGS measured with 2 hydraulic dynamometers (Jamar and Smedley) and 2 electric dynamometers (Nottingham Electronic and Saehan) in 118 participants aged 45–74 years. Data were analyzed using Bland-Altman, comparing maximum grip strength and the mean difference between electronic and hydraulic dynamometers. They concluded that there was a 4−5 kg disparity between the 2 groups. There was no difference when comparing dynamometers belonging to the same groups, electric vs electric and hydraulic vs hydraulic (difference ± 0.3 kg and ± 0.2 kg, respectively). According to the results obtained in the Bland-Altman analysis, the limits of agreement varied depending on the pair of devices compared. They were narrower (−2.0 and 10.1 kg) when comparing Saehan and Jamar, but very wide (−10.6 and 20.5 kg) when comparing Nottingham Electronic and Smedley dynamometers. Our study reports similar results, as Jamar and Saehan are the 2 dynamometers with the narrowest limits of agreement (LOA), especially in the PG (−6.47 and 5.70 kg).

Another recent study consolidates data similar to the above-mentioned study. In this study, Sumru Savas et al.11 compared the Jamar dynamometer with 2 others, the Saehan and the Takei TKK 5401. A total of 110 participants were included, with a mean age of 69.4 (SD, 6.3 years). The difference between the dynamometers was measured using Bland-Altman plots and ICC. The ICC in this study between Jamar and Saehan was 0.905 (SD, 0.865−0.934), very similar data to those collected in our study. The difference was 0.98 kg (LOA, 1.62 to −0.34) for the Jamar vs. Saehan dynamometer, limits closer to the mean than in our study, but with the same conclusion: Saehan has little difference and good correlation with Jamar.

The functional dynamometry test in the elderly population has a great impact on the diagnosis of sarcopenia and low muscle strength. It is a predictor of an impaired quality of life, which is why HGS monitoring in the elderly population is more important, as Guerra et al. 200912 did in 55 patients, with a mean age of 79.2 years (SD, 7.2). They used several dynamometers: Smedlay, Sammons Preston Rolyan, Bulb Eisenhut, and Jamar (reference). This study included data on nutritional status using the Mini Nutritional Assessment (MNA), a validated tool for evaluating and detecting nutritional status. A total of 13 elderly individuals (23.6%) were identified as at nutritional risk and 2 (3.6%) with malnutrition. They compared the 4 above-mentioned dynamometers and obtained a very wide range of measurements, data that coincide with our study. In this study, they corrected the range of data by subtracting the lowest value of the 3 consecutive measurements from the highest. Therefore, the association of the measurements approached 1, and they found an association between the measurements obtained by the Jamar device and the other instruments evaluated (r for Smedlay = 0.83; p < 0.001; r for Sammons Preston = 0.91; p < 0.001 and r for Eisenhut = 0.77; p < 0.001). Despite the high correlation coefficients, statistically significant differences were found between the mean HGS values evaluated with Jamar (24.8 ± 10.7 kg) and each of the other dynamometers: Smedlay (22.2 ± 8.6 kg), Eisenhut (14.5 ± 5.7 kg), and Sammons Preston (12.3 ± 6.3 kg). If we focus on the Bland-Altman analysis, between Jamar and Smedlay they found a mean difference of 3.2 kg (LOA, −6.3–12.6 kg).

The lack of consensus among measurement instruments may be due to 3 factors that differ between devices: the ergonomics of the grip handles, the transmission mechanism for measuring grip strength (hydraulic and mechanical), and the weight of the equipment. Regarding the ergonomics of different dynamometers, observing the difference in grip and hand position and the weight of the equipment, participants in the study called "Agreement-conformity between Camry and Jamar handheld dynamometers in adults," conducted by Muñoz et al. in 2018,13 stated that Jamar was heavier and gave other observations about the ease of holding the dynamometers. However, these observations were not measured, and therefore, it cannot be concluded that these variables are the cause of the difference in associations.

Certain strengths and limitations were considered. Regarding strengths, the study location at Hospital Universitario de León was considered, which includes the ability to obtain a population representation at a specific time, allowing for efficient identification of prevalences and associations between variables. Additionally, the inclusion of a control group of health care professionals younger than 40 years provided a solid basis for meaningful comparisons.

However, limitations are also important. The lack of randomization can introduce selection biases, and the cross-sectional nature of the study prevents observing changes over time, which limits the ability to establish causality.

Conclusions

Currently, many dynamometers are available on the market for measuring strength and evaluating patients to obtain more information about the impact of low muscle mass or the risk of sarcopenia. In our study, the dynamometers compared to the Jamar (considered the gold standard) showed an excellent correlation; however, both the bias and the dispersion of the measurements were significant, especially in the PG.

Although data indicate that dynamometers cannot be used interchangeably because the results do not have adequate agreement, if we consider the good correlation, as long as the same dynamometer and the same measurement method are used in all tests performed, the data can be validated in patient follow-up. However, values should be interpreted cautiously when using cut-off points in diagnosis obtained with another type of dynamometer.

This study does not consider differences in dynamometer ergonomics and weight. Regarding these variables and given the differences in associations presented by our study, collecting data on ergonomic and weight differences could be a good objective line of research to study whether there is an influence on HGS with respect to these parameters and the compared dynamometers.

Main findings

  • •

    Comparison with Jamar: Dynamometers compared to the Jamar (reference device) showed excellent correlation.

  • •

    Bias and Dispersion: Despite good correlation, there was significant bias and dispersion in measurements, especially in the patient group.

  • •

    Consistent Use: Dynamometers cannot be used interchangeably due to a lack of adequate agreement in results.

  • •

    Data Validation: If the same dynamometer and measurement method are used in all tests, data can be validated for patient follow-up, although other dynamometers should not be used interchangeably.

Declaration of competing interest

None declared.

References
[1]
R.W. Bohannon, K.L. Schaubert.
Test-retest reliability of grip-strength measures obtained over a 12-week interval from community-dwelling elders.
J Hand Ther., 18 (2005), pp. 426-428
[2]
M. León-Sanz, M. Brosa, M. Planas, A. García-de-Lorenzo, S. Celaya-Pérez, J.Á. Hernández, Predyces Group Researchers.
PREDyCES study: the cost of hospital malnutrition in Spain.
Nutrition (Burbank, Los Angeles County, Calif.), 31 (2015), pp. 1096-1102
[3]
A.J. Cruz-Jentoft, G. Bahat, J. Bauer, Y. Boirie, O. Bruyère, T. Cederholm, et al.
Sarcopenia: revised European consensus on definition and diagnosis.
Age Ageing, 48 (2019), pp. 601
[4]
N. Massy-Westropp, W. Rankin, M. Ahern, J. Krishnan, T.C. Hearn.
Measuring grip strength in normal adults: reference ranges and a comparison of electronic and hydraulic instruments.
J Hand Surg Am., 29 (2004), pp. 514-519
[5]
O. Shechtman, L. Gestewitz, C. Kimble.
Reliability and validity of the DynEx dynamometer.
J Hand Ther., 18 (2005), pp. 339-347
[6]
R.S. Guerra, T.F. Amaral.
Comparison of hand dynamometers in elderly people.
J Nutr Health Aging., 13 (2009), pp. 907-912
[7]
Y.C. Ha, S.C. Hwang, S.Y. Song, C. Lee, K.S. Park, J.I. Yoo.
Hand grip strength measurement in different epidemiologic studies using various methods for diagnosis of sarcopenia: a systematic review.
Eur Geriatr Med., 9 (2018), pp. 277-288
[8]
L. Huang, Y. Liu, T. Lin, L. Hou, Q. Song, N. Ge, et al.
Reliability and validity of two hand dynamometers when used by community-dwelling adults aged over 50 years.
BMC Geriatr., 22 (2022), pp. 580
[9]
M.J. Benton, J.M. Spicher, A.L. Silva-Smith.
Validity and reliability of handgrip dynamometry in older adults: a comparison of two widely used dynamometers.
[10]
C. Lessof, R. Cooper, A. Wong, et al.
Comparison of devices used to measure blood pressure, grip strength and lung function: a randomised cross-over study.
[11]
S. Savas, A. Kilavuz, F.Ö Kayhan Koçak, S. Cavdar.
Comparison of grip strength measurements by widely used three dynamometers in outpatients aged 60 years and over.
[12]
R.S. Guerra, T.F. Amaral.
Comparison of hand dynamometers in elderly people.
J Nutr Health Aging., 13 (2009), pp. 907-912
[13]
G.A.D. Muñoz, P.C. Martínez, V.C. Malagón.
Concordancia-conformidad entre los dinamómetros de mano Camry y Jamar en adultos.
Rev Nutr Clín Metabol, 1 (2018), pp. 35-41