Diferencias en la infraestimación de los volúmenes del ventrículo izquierdo entre 2 sistemas de ecocardiografía 3D. Comparación con resonancia magnética
Differences in the underestimation of left ventricular volumes between two 3-D Echography systems: A comparison with magnetic resonance
Visitas
1350
Juan Acosta Martínez
Autor para correspondencia
juan.acostamartinez@yahoo.es

Autor para correspondencia.
, José López Haldón, Justo Rodríguez Rodríguez, Israel Valverde, Maria Pilar Serrano, Nieves Romero, Silvia Gómez, Angel Martínez Martínez
Hospital Universitario Virgen del Rocío, Sevilla, España
Este artículo ha recibido
1350
Visitas
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (1)
Tablas (1)
Tabla 1. Enfermedad cardíaca que indicó el estudio de cardiorresonancia magnética
Resumen
Introducción

La ecocardiografía 3D (E3D) es más precisa que la técnica 2D en la valoración de los volúmenes y la fracción de eyección (FE) del ventrículo izquierdo (VI). Este estudio compara, con respecto a cardiorresonancia magnética (cRM), 2sistemas de E3D: E3D en un solo latido con detección automática de bordes endocárdicos (Siemens SC2000;A) frente a E3D con adquisición de varios latidos y detección semiautomática de bordes endocárdicos (Philips iE33;B).

Métodos

A 22 pacientes remitidos para estudio mediante cRM se les realizó E3D con sistema de adquisición en un solo latido y detección automática de bordes (A) y con adquisición en varios latidos y detección semiautomática de bordes (B).

Resultados

Los volúmenes y la FE medios por cRM fueron: VTD: 211,07±85,4ml; VTS: 119,2±89,25ml; FE: 48,67±20,07%. La correlación entre la cRM y los 2sistemas de E3D fue excelente para el VTD (rhoA 0,912; rhoB 0,907; p<0,0001), VTS (rhoA 0,877; rhoB 0,945; p<0,0001) y FE (rhoA 0,974; rhoB 0,951; p<0,0001). El análisis mediante Bland-Altman reveló infraestimación de los volúmenes del VI con ambos sitemas de E3D, siendo esta menor con el sistemaA queB (–40±50,2 vs –82,2±40,9ml, respectivamente, para VTD; –16±44,4 vs –43±43,5ml, respectivamente, para VTS; p=0,001). No hubo diferencias en la medida de la FE entre las 3técnicas de imagen.

Conclusión

La E3D infraestima los volúmenes del VI. En nuestra serie esta infraestimación fue menor con el sistema de adquisición en único latido y detección automática de bordes.

Palabras clave:
Ecocardiografía tridimensional
Volúmenes
Ventrículo izquierdo
Abstract
Introduction

3-D echocardiography (3DE) is superior to the 2-D technique in assessing left ventricular (LV) volumes and ejection fraction (EF). This study compares two 3DE systems with cardiac magnetic resonance imaging (cMRI): single-beat 3DE with automated endocardiac border detection (Siemens SC2000;A) and 3DE with multi-beat acquisition and semi-automated endocardiac borders detection (Philips iE33;B).

Methods

A 3DE with single-beat acquisition and automated border detection (A) and 3DE with multi-beat acquisition and semi-automated border detection (B) was performed on 22patients referred for a study using cMRI.

Results

The mean volumes and EF measured by cMRI were: DTV: 211.07±85.4mL; STV: 119.2±89.25mL; EF: 48.67±20.07%. The correlation between cMRI and the two 3DE systems was excellent for the DTV (rhoA 0.912; rhoB 0.907; P<.0001), STV (rhoA 0.877; rhoB 0.945; P<.0001) and EF (rhoA 0.974; rhoB 0.951; P<.0001). The Bland-Altman analysis showed an underestimation of the LV volumes with both 3DE systems, with the difference being lower in systemA than in systemB (–40±50.2 vs –82.2±40.9mL, respectively for DTV; –16±44.4 vs –43±43.5mL, respectively for STV; P=.001). There were no differences in the measurement of EF between the 3imaging techniques.

Conclusion

3DE underestimates LV volumes. In our series, this underestimate was lower in the system with single-beat acquisition and automated border detection.

Keywords:
Three-dimensional echocardiography
Volumes
Left ventricle

Artículo

Opciones para acceder a los textos completos de la publicación Cardiocore
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Cardiocore

Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
Herramientas
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos

es en pt
Política de cookies Cookies policy Política de cookies
Utilizamos cookies propias y de terceros para mejorar nuestros servicios y mostrarle publicidad relacionada con sus preferencias mediante el análisis de sus hábitos de navegación. Si continua navegando, consideramos que acepta su uso. Puede cambiar la configuración u obtener más información aquí. To improve our services and products, we use "cookies" (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here. Utilizamos cookies próprios e de terceiros para melhorar nossos serviços e mostrar publicidade relacionada às suas preferências, analisando seus hábitos de navegação. Se continuar a navegar, consideramos que aceita o seu uso. Você pode alterar a configuração ou obter mais informações aqui.