Original Article
Open Access
Percutaneous occlusion of the left atrial appendage with AMPLATZER® Cardiac Plug for the prevention of thromboembolic events in chronic atrial fibrillation
Visits
3411
Francisco Chamié
Corresponding author
fchamie@pobox.com

Corresponding author: Rua Real Grandeza, 108, salas 223-224, Botafogo, CEP: 22281-034, Rio de Janeiro, RJ, Brazil.
, Enio Guérios, Valério Fuks, Angélica Bösiger, Marcio Carvalho, João Otávio de Queiroz Araujo
Hospital Federal dos Servidores do Estado, Rio de Janeiro, RJ, Brazil
This item has received

Under a Creative Commons license
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (7)
Show moreShow less
Tables (2)
Table 1. Clinical data of patients submitted to left atrial appendage occlusion.
Table 2. Characteristics of occluded appendages and devices used.
Show moreShow less
ABSTRACT
Background

Atrial fibrillation (AF) increases the risk of thromboembolic events caused by emboli originating in the left atrial appendage (LAA). Mechanical methods for LAA occlusion have been developed as an alternative to oral anticoagulation. The aim of this study was to present an initial experience with the AMPLATZER® Cardiac Plug.

Methods

Patients with permanent or paroxysmal AF and with contraindications or complications of oral anticoagulation were included. Patients with LAA anatomy and measures compatible with the occluder, and without thrombi, were selected through transesophageal echocardiography.

Results

A total of 14 procedures were performed in 13 patients (5M:8F), with mean age of 66.7 years. Significant bleeding and previous strokes were found in 69.2% and 53.8%, respectively. AF was permanent in 84.6% and paroxysmal in the remainder. The mean diameters of the ostium and the landing zone were 23.9mm and 20.8mm, respectively. Bilobulated LAA was observed in 76.9%. Procedures were possible in all cases. Sixteen devices were used in 13 patients, a ratio of 1.2:1, and only one patient required a second device for LAA occlusion. The mean follow-up was 12.2 months. All LAA remain closed, with no residual defect to date. There was only one late death, unrelated to the procedure.

Conclusions

LAA occlusion using the AMPLATZER® Cardiac Plug device was shown to be safe and effective in this small series of patients. The initial results are encouraging and indicate the transcatheter closure of the LAA as an alternative to oral anticoagulation therapy in selected patients.

Keywords:
Thromboembolism
Stroke
Atrial fibrillation
RESUMO
Introdução

A fibrilação atrial (FA) aumenta o risco de eventos tromboembólicos por êmbolos originados em apêndice atrial esquerdo (AAE). Métodos mecânicos para a oclusão do AAE foram desenvolvidos como alternativa à anticoagulação oral. O objetivo deste trabalho foi apresentar uma experiência inicial com o AMPLATZER® Cardiac Plug.

Métodos

Incluímos pacientes com FA permanente ou paroxística, que apresentavam contraindicações ou complicações derivadas da anticoagulação oral. Pacientes com anatomia e medidas do AAE compatíveis com o oclusor, e sem trombos foram selecionados por meio de ecocardiograma transesofágico.

Resultados

Foram realizados 14 procedimentos em 13 pacientes (5M:8F), com média de idade de 66,7 anos. Sangramento significativo e acidentes vasculares cerebrais prévios foram encontrados em 69,2% e em 53,8%, respectivamente. A FA era permanente em 84,6% e paroxística no restante da amostra. Os diâmetros médio do óstio e da zona alvo mediram 23,9mm e 20,8mm, respectivamente. AAE bilobulados foram observados em 76,9%. Os procedimentos foram possíveis em todos os casos. Dezesseis dispositivos foram usados em 13 pacientes, numa razão de 1,2:1, e apenas 1 paciente precisou de um segundo dispositivo para oclusão do AAE. O tempo médio de acompanhamento foi de 12,2 meses. Todos os AAE permanecem fechados e sem defeito residual até o momento. Houve apenas um óbito tardio não relacionado ao procedimento.

Conclusões

A oclusão do AAE com o dispositivo de AMPLATZER® Cardiac Plug mostrou ser segura e eficaz nesta pequena série de pacientes. Os resultados iniciais são encorajadores e apontam para o fechamento transcateter do AAE como alternativa para a anticoagulação oral em pacientes selecionados.

Palavras-chave:
Tromboembolia
Acidente vascular cerebral
Fibrilação atrial
Full Text
Introduction

Atrial fibrillation (AF) of non-valvular origin is one of the most common type of arrhythmia, affecting 1 to 2% of the overall adult population.1 It is present in up to 14% of patients older than 65 years and its incidence doubles every decade.2 Being associated with a high risk of cardioembolic events, particularly stroke, AF accounts for approximately 15% of all ischemic strokes.3 The risk of stroke in patients with AF increases up to five times when compared to patients in sinus rhythm.4

The prevention of embolic phenomena of AF is traditionally achieved through the continued use of oral anticoagulants (OAC). The most often used OAC are vitamin K antagonists, which are effective, reducing the incidence of stroke in up to 60% and death in up to 25% of patients that remain within the therapeutic range.5 The continued use of these drugs brings several difficulties and inconveniences to patients, resulting in a large number of untreated individuals. The main problems reported are the risk of bleeding or previous episodes of significant bleeding in individuals with predisposing conditions; extreme frailty; the low rate of adherence to treatment; harmful drug interactions; the oscillation of therapeutic levels of the drug; and unwanted side effects.6–9

The knowledge that more than 90% of emboli in AF originate from thrombi in the left atrial appendage (LAA) led to the development of options for mechanical obliteration of this structure as an alternative therapy to OAC to prevent cerebral thromboembolic phenomena.10–12

In this article, the authors report their initial experience with LAA occlusion in a single center, using the first dedicated prosthesis developed for this purpose and approved for clinical use in Brazil.

Methods

Records of all patients referred for percutaneous closure of the LAA in the Interventional Cardiology Department of Structural and Congenital Defects of Hospital Federal dos Servidores do Estado in the city of Rio de Janeiro (RJ) were retrospectively analyzed.

Patients with permanent or paroxysmal AF and who had contraindications or complications caused by the continued use of OAC were selected. Selection for the procedure was attained by transesophageal echocardiography (TEE), which included patients whose atrial appendages had anatomical characteristics and diameters compatible with the standard occluder used in this service (12.6mm to 28.5mm)13 and who did not have thrombi inside the LAA.

The prosthesis used was the AMPLATZER® Cardiac Plug (ACP, AGA Medical Corp., Minneapolis, USA), which is a self-expanding device made of nitinol, lined with polyester fabric. It consists of a cylindrical lobe available in diameters of 16 to 30mm, with 2-mm increments, which corresponds to the nominal size of the prosthesis. Attached to it by a flexible connector pin is a disc; the disc is 4mm larger than the lobe in 16 to 22-mm prostheses, and 6mm larger in 24 to 30-mm prostheses.

To increase the safety of the procedure and reduce the risk of device displacement, six pairs of thin stabilizing hooks are fastened to the lobe, identified by radiopaque markers, which help to secure the lobe to the LAA body in the landing zone (Fig. 1).

Figure 1.

Occluder details. Note the disc at the bottom of the picture and, above it, the lobe, in which six pairs of locking hooks are displayed.

(0.1MB).

All procedures were performed under general anesthesia and tracheal intubation, after a minimum fasting period of 8hours, under fluoroscopic and transesophageal echocardiography guidance in the interventional cardiology laboratory.

Unfractionated heparin was administered at doses of 5,000 to 10,000 IU after obtaining transseptal access. Supplemental doses of 2,500 to 5,000 IU were administered every 30minutes when the procedures lasted longer than 1 hour. Antimicrobial prophylaxis with 2g of intravenous cefazolin was routinely administered. The patients were submitted to right and left heart catheterization through femoral vein puncture. Left atrial access was obtained by transseptal puncture with the standard technique using a Brockenbrough needle.

With the aid of a calibrated 5 F pigtail catheter, the LAA was catheterized and injections were performed in the right anterior oblique (RAO) view, with cranial and caudal angulation, to determine the anatomic type and to obtain measures of the ostium and the landing zone of the occluder device's lobe. The measures and anatomy were compared with those obtained with the TEE, with the largest measurements utilized to select the prosthesis to be implanted (Fig. 2).

Figure 2.

Correspondence between angiographic and echocardiographic images. In A, the left atrial appendage seen in the cranial-right anterior oblique view shows similarity with the echocardiographic image obtained with the transducer at 55° in B. In C, appendage angiography in the right anterior oblique view with caudal angulation shows a better depiction of the more terminal trabecular portion of the left atrial appendage, which is also seen in D in the echocardiographic image at 132°.

(0.12MB).

The pigtail catheter was removed and replaced by a long guide wire (260cm) with a curved tip (J), rigid or super rigid, inside the LAA. Over it, a long, double-curve AGA sheath was introduced, with a diameter compatible with the chosen occluder. The sheath was positioned as coaxially as possible to the LAA axis and the position was assessed using manual contrast injections, through the hemostatic valve. The chosen device was introduced through the sheath, previously loaded in the delivery system, whose lobe diameter was 2 to 4mm larger than the largest measure obtained from the implantation landing zone. The device lobe was ideally implanted maintaining two-thirds of the lobe in a distal position to the left circumflex coronary artery location, visualized by TEE. Subsequently, while maintaining slight traction on the delivery cable, the sheath was retracted, externalizing the disk in order to occlude the LAA ostium. The tension on the cable was maintained for a few minutes (tug test) to ensure that the device was well adhered to the LAA body, minimizing the risk of embolization or displacement. The evaluated signs of good positioning were the degree of compression of the occluder lobe, separation between the disc and the lobe, the concave shape of the disc and its position at the ostium entrance, and lobe positioning with at least two-thirds within the left circumflex coronary artery (Fig. 3).

Figure 3.

Three-dimensional transesophageal echocardiography image showing some of the signs of adequate device positioning, such as lobe compression and the separation between the lobe and the disc, which shows a concave configuration, occluding the ostium of the left atrial appendage.

(0.07MB).

New injections were performed through the delivery system, to confirm the correct positioning of the device with complete occlusion of the LAA ostium (Fig. 4). After it was considered to be satisfactory, the prosthesis was released from the delivery cable by unscrewing it from the delivery system. Immediately after the implantation, the absence of periprosthetic residual flow and pericardial effusion was verified through the TEE.

Figure 4.

Angiography performed through the sheath system showing the disk inside the left atrial appendage and complete obliteration of the ostium by the prosthesis disc still attached to the delivery system, which would allow its repositioning if necessary.

(0.11MB).

Patients were awakened in the interventional laboratory and taken to the intensive care unit for postoperative care. Another dose of 1g of cefazolin was administered 6hours after the procedure. Patients were discharged on the day after the implantation, after being submitted to a control transthoracic echocardiogram (TTE).

At the follow-up, all patients were instructed to use dual antiplatelet therapy with 200mg of acetylsalicylic acid associated with 75mg of clopidogrel bisulfate. Medical evaluations followed by echocardiographic assessment were performed after 1 and 3 months, when clopidogrel was discontinued. After 6 months, control TEE was performed and patients maintained the continuous use of acetylsalicylic acid.

Results

From March 2014 to August 2015, 14 percutaneous occlusions of the LAA were performed in 13 patients. Eight patients were females. Age ranged from 46 to 83 years (66.7 ± 11.8 years).

Significant bleeding during OAC use was found in nine (69.2%) patients. Gastrointestinal bleeding was identified in six patients, and a retinal, genitourinary and intracranial bleeding in the remainder.

Seven (53.8%) patients had suffered a previous stroke. Most patients (84.6%) had permanent AF; in the others, the arrhythmia was paroxysmal. Clinical data and the main comorbidities found are described in Table 1.

Table 1.

Clinical data of patients submitted to left atrial appendage occlusion.

ID  Age (years)  Indication  Comorbidities  CHA2DS2-VASc  HAS-BLED  Atrial fibrillation 
AFS  71  Gastrointestinal bleeding + stroke  SAH + DM 2  Paroxysmal 
JSFF  78  Gastrointestinal bleeding  Permanent 
GC  46  Stroke  SAH + DCM  Paroxysmal 
SMGQ  66  Retinal bleeding  SAH + DAC + PFO  Permanent 
ABB  80  Gastrointestinal bleeding + stroke  SAH + DM2  Permanent 
MPG  53  Labile INR  SAH + CAD  Permanent 
RSRS  76  Gastrointestinal bleeding  SAH + DM2  Permanent 
GAS  65  Stroke  SAH + DCM  Permanent 
HBS  58  Genitourinary bleeding  DCM  Permanent 
10  MJML  71  Stroke  SAH + CAD  Permanent 
11  AIF  52  IC bleeding + stroke  SAH + DCM + hypothyroidism  Permanent 
12  MJB  83  Gastrointestinal bleeding + stroke  SAH + DM2 + DCM  Permanent 
13  GSA  76  Gastrointestinal bleeding  SAH + DM2 + CAD + DCM  Permanent 

ID: patient identification; SAH: systemic arterial hypertension; DM2: type 2 diabetes mellitus; DCM: dilated cardiomyopathy; CAD: coronary artery disease; PFO: patent foramen ovale; INR: international normalized ratio.

Regarding the shape, bilobulated appendages were present in 76.9% of cases, whereas the others were single-lobulated. Ostial diameter ranged from 13 to 30mm (23.9 ± 5.4mm) and the landing zone from 11 to 28mm (20.8 ± 5.1mm).

The procedure was successful in all cases. Sixteen devices were used for occlusion in the 13 patients, at a ratio of 1.2 device per patient. The other AMPLATZER® PFO Occluder (PFO) prosthesis of 25mm was used to occlude the foramen ovale of a patient (SMGQ, case 4) who did not require transseptal puncture and had her LAA occluded with ACP through the foramen ovale (Fig. 5).

Figure 5.

Images from case 4. The AMPLATZER® 24-mm Cardiac Plug device is seen occluding the left atrial appendage and the AMPLATZER® 25-mm PFO Occluder device is positioned through the patient's foramen ovale. For the implant of the AMPLATZER® PFO Occluder device, the procedure utilized the same double curve sheath that was used for the implant of the AMPLATZER® Cardiac Plug device.

(0.09MB).

Only one patient required more than one prosthesis to occlude the LAA (ABB, case 5). He had his occlusion procedure performed as a live case during Rio.Interv 2014. At the time, a proximal accessory lobe was occluded in the anterosuperior position with a 16-mm ACP, and a 25-mm PFO prosthesis was used for the remainder of the appendage body, with immediate success. After 2 days, the PFO prosthesis migrated to the transverse aorta, from which it was percutaneously removed with a loop snare catheter. As the main body of the appendage was uncovered after embolization of the PFO prosthesis, a second procedure was offered and carried out 5 months after the index procedure. Initially, an attempt was made to occlude the LAA with a 30-mm ACP, which showed to be oversized and was finally replaced by a 28-mm ACP, which adequately occluded the appendage (Fig. 6). In the remaining patients, the LAA was occluded with a single device. Fifteen ACP and two PFO prostheses were used for LAA occlusion (Table 2).

Figure 6.

Images from case 5. In A, the accessory lobe occluded with the AMPLATZER® 16-mm Cardiac Plug and the appendage body occluded with AMPLATZER® 25-mm PFO Occluder can be observed. In B, the AMPLATZER® PFO Occluder is embolized into the aortic arch, being captured by the loop catheter. In C, at the end of the final procedure, the accessory lobe is completely occluded by AMPLATZER® 16-mm Cardiac Plug and the body of the left atrial appendage by the second AMPLATZER® 28-mm Cardiac Plug device. In D, echocardiographic control image after the procedure, showing the ostium of the left atrial appendage occluded by AMPLATZER® 28-mm Cardiac Plug device. Observe the increase in echocardiographic density in the left atrial appendage corresponding to AMPLATZER® 16-mm Cardiac Plug device, which is barely visible in this image.

(0.14MB).
Table 2.

Characteristics of occluded appendages and devices used.

ID  Shape  Ostium (mm)  Landing zone (mm)  Devices used 
AFS  Bilobulated  24  22  ACP 26 
JSFF  Single-lobulated  28  26  ACP 30 
GC  Bilobulated  28  28  ACP 30 
SMGQ  Bilobulated  24  17  ACP 24 + PFO 25 
ABB  Bilobulated  26  26  ACP 16 + PFO 25 + ACP 30 + ACP 28 
MPG  Bilobulated  13  11  ACP 16 
RSRS  Bilobulated  19  18  ACP 22 
GAS  Bilobulated  20  18  ACP 24 
HBS  Single-lobulated  30  22  ACP 26 
10  MJML  Single-lobulated  16  15  ACP 18 
11  AIF  Bilobulated  30  24  ACP 28 
12  MJB  Bilobulated  28  26  ACP 30 
13  GSA  Bilobulated  25  18  ACP 20 

ID: patient identification; ACP: AMPLATZER® Cardiac Plug; PFO: patent foramen ovale.

There was a single periprocedural complication (MPG, case 6). The patient had a narrow and shallow appendage, which was perforated by the hydrophilic guide at the beginning of the procedure. She had pericardial effusion, which was promptly drained at the interventional cardiology laboratory. After hemodynamic stabilization of the patient, the procedure continued and the LAA was occluded with a 16-mm ACP (Fig. 7). The effusion was contained after appendage occlusion and the patient was transferred to the intensive care unit in good condition for post-procedure care. At the follow-up, she had persistent pericarditis treated with colchicine and corticosteroids; she achieved remission after 30 days and has remained asymptomatic since then.

Figure 7.

Images from case 6. In A, angiography performed through the double curve sheath, showing the small perforation at the tip of the left atrial appendage with contrast extravasation into the pericardial cavity. In B, after hemodynamic stabilization of the patient by draining the pericardial effusion with a pigtail catheter, the left atrial appendage totally occluded by the AMPLATZER® 16-mm Cardiac Plug device can be observed.

(0.13MB).

Follow-up was obtained in all cases. The mean time was 12.2 months (5 to 19 months) and only one patient had not yet reached the 6-month follow-up period. All atrial appendages remained closed, and no patient had periprosthetic leaks nor thromboembolic events to date.

There was one late death, which was unrelated to the procedure. The patient (GC, case 3) had severe dilated cardiomyopathy and died of refractory heart failure 11 months after the procedure. The prosthesis remained correctly positioned and there was no periprosthetic regurgitation.

Discussion

The first, fundamental step in the occlusion procedure is obtaining access to the left atrium. Typically, this is accomplished through transseptal puncture with a Brockenbrough needle.14 To access and manipulate the long sheath coaxially inside the LAA, it is important that the transseptal needle cross the atrial septum in the posteroinferior position.13 Visualization by TEE allows the interventionist to perform the transseptal puncture precisely in the desired place, favoring procedural success.15 Additionally, gaining access to the left atrium via atrial septal defects has been described for LAA occlusion. The transseptal puncture is avoided and the technique does not prevent the completion of the procedures in most cases.16,17

The anatomy and the spatial location of the LAA are extremely variable, resulting in unique descriptions (cauliflower, cactus, chicken wing, windsock),13 which were not used in this study. The anatomical analysis of the LAA, in this study, was performed through transesophageal echocardiography and angiography. A correlation was observed between the echocardiographic images obtained with the transducer positioned between zero and 60° and the angiograms performed in RAO view with cranial angulation, whereas the images obtained with the transducer positioned between 90° and 130° were better correlated with angiograms obtained in RAO view with caudal angulation, which more precisely delineated the terminal part of the trabecular portion of the appendages. In the limited experience reported in this article, the RAO view was preferably used with caudal angulation for device implantation, as it has demonstrated the highest measured values, and the latter guided the choice of prostheses used.

The LAA, together with its great anatomical variability, is an extremely delicate and friable structure, which may explain the incidence of pericardial effusion observed in the occlusion procedures, ranging from 1.2 to 3.7%.18–21

Extreme care should be exercised when manipulating it and some basic rules need to be followed, such as using only round-tip guide wires (J) and striving to maintain the long delivery sheath as coaxial as possible to the long axis of the appendage, away from its walls. Care must be taken at all times during the procedure with the LAA borders, avoiding at all costs reaching its extremity with the dilator or long sheath. The learning curve of this procedure is significant, and it has been shown that the number of procedural complications decreases as the team becomes familiar with the occlusion technique.22,23

The ACP consists of a lobe and a disk connected by a flexible pin that articulates and allows angulation between them. The lobe, which is implanted in the LAA body, has 12 small hooks for device stabilization. It is of utmost importance that, to prevent embolization of the prosthesis, the lobe undergoes some degree of compression, to promote contact between the hooks and the LAA walls. For this purpose, it is important to use the largest measurements obtained, choosing devices that are 4 to 6mm larger than the landing zone. However, the use of prosthetic devices much larger than the implantation region causes device migration, as seen in case 5. An excessively oversized device deforms the lobe, modifying the positioning of the hooks, retracting them and causing loss of their contact with the LAA wall. The same occurs if the selected device is too small, smaller than the landing zone, a situation in which also there is no contact of the hooks with the LAA wall.

The implantation of the prosthesis should promote appendage body occlusion, isolating its entire trabecular portion, whose uneven surface creates conditions for the formation of thrombi found in AF. The presence of the disc, positioned aiming to obliterate the LAA ostium, acts as a second occluder element, increasing procedural safety. It is important for the disk to maintain a concave shape and to adapt perfectly to the ostium, preventing spaces that can be a source of thrombi or even the origin of periprosthetic regurgitation.

It is presumed that periprosthetic residual flow may be the cause of new cerebral ischemic strokes and, therefore, should be prevented, although there is some controversy in the literature.24,25 Currently, residual flows with diameter ≥ 5mm are considered as significant,26 which, when present, should be occluded with a second device. The use of more than one device for LAA occlusion has been successfully performed and appears to be associated with favorable results in patient follow-up.27

Conclusions

The occlusion of the left atrial appendage in this small number of cases showed to be safe and effective, in spite of the complexity and comorbidities of the patients included in this study.

The initial results are encouraging and, despite the short-term follow-up of these patients, they confirm that percutaneous occlusion of the left atrial appendage can become a valid alternative to oral anticoagulation in patients who cannot use it.

Funding source

None declared.

Conflicts of interest

Enio Guérios is a consultant and proctor of St. Jude Medical.

References
[1]
P.A. Wolf, T.R. Dawber, H.E. Thomas Jr., W.B. Kannel.
Epidemiologic assessment of chronic atrial fibrillation and risk of stroke: the Framingham study.
Neurology., 28 (1978), pp. 973-977
[2]
W.M. Feinberg, J.L. Blackshear, A. Laupacis, R. Kronmal, R.G. Hart.
Prevalence, age distribution, and gender of patients with atrial fibrillation. Analysis and implications.
Archives of internal medicine, 155 (1995), pp. 469-473
[3]
P. Sandercock, J. Bamford, M. Dennis, J. Burn, J. Slattery, L. Jones, et al.
Atrial fibrillation and stroke: prevalence in different types of stroke and influence on early and long term prognosis (Oxfordshire community stroke project).
BMJ, 305 (1992), pp. 1460-1465
[4]
P.A. Wolf, R.D. Abbott, W.B. Kannel.
Atrial fibrillation as an independent risk factor for stroke: the Framingham Study.
Stroke., 22 (1991), pp. 983-988
[5]
R.G. Hart, L.A. Pearce, M.I. Aguilar.
Meta-analysis: antithrombotic therapy to prevent stroke in patients who have nonvalvular atrial fibrillation.
Ann Intern Med., 146 (2007), pp. 857-867
[6]
S. Gafoor, H. Sievert, P. Matic.
Close the appendage, open the future.
[7]
R.S. Stafford, D.E. Singer.
Recent national patterns of warfarin use in atrial fibrillation.
Circulation., 97 (1998), pp. 1231-1233
[8]
L. Wallentin, S. Yusuf, M.D. Ezekowitz, M. Alings, M. Flather, M.G. Franzosi, et al.
RE-LY investigators. Efficacy and safety of dabigatran compared with warfarin at different levels of international normalised ratio control for stroke prevention in atrial fibrillation: an analysis of the RE-LY trial.
Lancet., 376 (2010), pp. 975-983
[9]
E.M. Hylek, S.J. Skates, M.A. Sheehan, D.E. Singer.
An analysis of the lowest effective intensity of prophylactic anticoagulation for patients with nonrheumatic atrial fibrillation.
N Engl J Med., 335 (1996), pp. 540-546
[10]
J.L. Blackshear, J.A. Odell.
Appendage obliteration to reduce stroke in cardiac surgical patients with atrial fibrillation.
Ann Thorac Surg., 61 (1996), pp. 755-759
[11]
W.J. Manning, D.I. Silverman, C.A. Waksmonski, P. Oettgen, P.S. Douglas.
Prevalence of residual left atrial thrombi among patients with acute thromboembolism and newly recognized atrial fibrillation.
Arch Intern Med., 155 (1995), pp. 2193-2198
[12]
M.F. Stoddard, P.R. Dawkins, C.R. Prince, N.M. Ammash.
Left atrial appendage thrombus is not uncommon in patients with acute atrial fibrillation and a recent embolic event: a transesophageal echocardiographic study.
J Am Coll Cardiol., 25 (1995), pp. 452-459
[13]
J.A. Cabrera, F. Saremi, D. Sánchez-Quintana.
Left atrial appendage: anatomy and imaging landmarks pertinent to percutaneous transcatheter occlusion.
Heart., 100 (2014), pp. 1636-1650
[14]
P.G. Nixon, H. Ikram.
Left heart catheterization with special reference of the transeptal method.
Br Heart J., 28 (1965), pp. 835-841
[15]
I.C. Ho, P. Neuzil, T. Mraz, Z. Beldova, D. Gross, P. Formanek, et al.
Use of intracardiac echocardiography to guide implantation of a left atrial appendage occlusion device (PLAATO).
Heart Rhythm., 4 (2007), pp. 567-571
[16]
F. Chamié, D. Chamié, L.C. Simões, E. Guérios, J.C. Tress.
Oclusão percutânea do apêndice atrial esquerdo e do forame oval patente no mesmo procedimento sem necessidade de punção transeptal.
Rev Bras Cardiol Invas., 22 (2014), pp. 382-385
[17]
D. Koermendy, F. Nietlispach, S. Shakir, S. Gloekler, P. Wenaweser, S. Windecker, et al.
Amplatzer left atrial appendage occlusion through a patent foramen ovale.
Catheter Cardiovasc Interv., 84 (2014), pp. 1190-1196
[18]
O.O. Alli, D.R. Homes Jr..
Left atrial appendage occlusion for stroke prevention.
Curr Probl Cardiol., 40 (2015), pp. 429-476
[19]
Y.Y. Lam, G.W. Yip, C.M. Yu, W.W. Chan, B.C. Cheng, B.P. Yan, et al.
Left atrial appendage closure with AMPLATZER cardiac plug for stroke prevention in atrial fibrillation: initial Asia-Pacific experience.
Catheter Cardiovasc Interv., 79 (2012), pp. 794-800
[20]
J.W. Park, A. Bethencourt, H. Sievert, G. Santoro, B. Meier, K. Walsh, et al.
Left atrial appendage closure with Amplatzer cardiac plug in atrial fibrillation: initial European experience.
Catheter Cardiovasc Interv., 77 (2011), pp. 700-706
[21]
A. Tzikas, S. Shakir, S. Gafoor, H. Omran, S. Berti, G. Santoro, et al.
Left atrial appendage occlusion for stroke prevention in atrial fibrillation: multicentre experience with the AMPLATZER Cardiac Plug.
EuroIntervention., 10 (2015),
pii: 20140825-01
[22]
D.R. Homes Jr., S. Kar, M.J. Price, B. Whisenant, H. Sievert, S.K. Doshi, et al.
Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial.
J Am Coll Cardiol., 64 (2014), pp. 1-12
Erratum in: J Am Coll Cardiol. 2014;64(11):1186.
[23]
V.Y. Reddy, D. Holmes, S.K. Doshi, P. Neuzil, S. Kar.
Safety of percutaneous left atrial appendage closure: results from the Watchman Left Atrial Appendage System for Embolic Protection in Patients with AF (PROTECT AF) clinical trial and the Continued Access Registry.
Circulation., 123 (2011), pp. 417-424
[24]
X. Freixa, A. Tzikas, A. Sobrino, J. Chan, A.J. Basmadjian, R. Ibrahim.
Left atrial appendage closure with the Amplatzer Cardiac Plug: impact of shape and device sizing on follow-up leaks.
Int J Cardiol., 168 (2013), pp. 1023-1027
[25]
E.S. Katz, T. Tsiamtsiouris, R.M. Applebaum, A. Schwartzbard, P.A. Tunick, I. Kronzon.
Surgical left atrial appendage ligation is frequently incomplete: a transesophageal echocardiograhic study.
J Am Coll Cardiol., 36 (2000), pp. 468-471
[26]
J. Wiebe, J. Franke, K. Lehn, I. Hofmann, L. Vaskelyte, S. Bertog, et al.
Percutaneous left atrial appendage closure with the Watchman device: long-term results up to 5 years.
JACC Cardiovasc Interv., 8 (2015), pp. 1915-1921
[27]
ÊE Guérios, S. Gloekler, M. Schmid, A. Khattab, F. Nietlispach, B. Meier.
Double device left atrial appendage closure.
EuroIntervention., 11 (2015), pp. 470-476

Peer review under the responsibility of Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista.

Copyright © 2015. Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista