Buscar en
Enfermedades Infecciosas y Microbiología Clínica (English Edition)
Toda la web
Inicio Enfermedades Infecciosas y Microbiología Clínica (English Edition) Microbial contamination of liposomal amphotericin B nebuliser devices in lung tr...
Journal Information
Vol. 37. Issue 3.
Pages 211-212 (March 2019)
Vol. 37. Issue 3.
Pages 211-212 (March 2019)
Scientific letter
Full text access
Microbial contamination of liposomal amphotericin B nebuliser devices in lung transplant patients
Contaminación microbiana de dispositivos de nebulización de anfotericina B liposomal en pacientes trasplantados pulmonares
Visits
1063
Miguel Fernández-Huertaa,
Corresponding author
miguel.fernandez@vhebron.net

Corresponding author.
, Roser Escobarb, Victor Monforteb, Virginia Rodrigueza
a Servicio de Microbiología, Hospital Universitario Vall d’Hebron, Universitat Autònoma de Barcelona, Barcelona, Spain
b Servicio de Pneumología, Hospital Universitario Vall d’Hebron, Universitat Autònoma de Barcelona, Barcelona, Spain
This item has received
Article information
Full Text
Bibliography
Download PDF
Statistics
Tables (1)
Table 1. Microorganisms isolated in the nebulisers.
Full Text

Microbiological Procedure number 61 of the Sociedad Española de Enfermedades Infecciosas y Microbiología Clínica (SEIMC) [Spanish Society of Infectious Diseases and Clinical Microbiology] entitled “Microbiological methods for monitoring the cleaning, disinfection and sterilisation of medical devices” focuses on reusable medical devices qualified as semi-critical or medium risk; those that come into contact with the mucous membranes without penetrating sterile tissues and which require at least high-level disinfection. With this in mind, we would like to present a study about microbial contamination of liposomal amphotericin B nebulising equipment in lung transplant patients.

Nebulisers are devices for administering a drug through the airways by aerosol delivery. In the context of lung transplantation, home-administered prophylaxis with nebulised amphotericin B is effectively used for the prevention of infection by Aspergillus spp.1 These devices are a potential source of infection in the event of microbial contamination.2 Very few studies have assessed the incidence of microbial contamination in nebulisers and the clinical repercussions,3 particularly in lung transplantation.4 Moreover, there is a lack of consensus on the standardisation of cleaning and disinfection procedures and microbiological monitoring of these devices.5 The aim of this study is to estimate the prevalence of microbial contamination in amphotericin B nebulising systems used by lung transplant patients, and to assess the impact on the patient's microbiology.

When the prophylaxis was starting, all patients were instructed on how to adequately clean and disinfect the nebuliser (washing with a brush, soap and water and immersion in 1% sodium hypochlorite solution). After the transplant, the patients had visits every month for the first year, every two months in the second year and then every three to four months. Sputum cultures were taken at these visits if the patient expectorated and fibreoptic bronchoscopy was also occasionally performed.

In 2012–2014, the nebulisation equipment of a cohort of lung transplant patients was sampled. We studied 75 cultures from 71 patients; 42 males, with a mean age of 49 (range: 5–64 years), and 29 females, with a mean age of 51 (22–66 years). Samples were taken from the pipette, the reservoir and tubing of the device. The median time between transplantation and collection of the sample was 666 days (range: 68–1470 days). The results of the cultures of respiratory samples collected in the three months before and after the sampling of the nebuliser were also reviewed.

Microbial contamination was detected in 53.3% of the nebulisers (Table 1). The contamination rate was significantly higher in the pipette and the reservoir (50.7%) than in the device tubing (11.5%) (p<0.005). Pseudomonas aeruginosa was not isolated in any of the nebulisers.

Table 1.

Microorganisms isolated in the nebulisers.

Nebulising devices studied (75)  Number  Percentage (95% CI) 
Negative  35  46.7% (35.1–58.6%) 
Positivea  40  53.3% (41.4–64.9%) 
Non-fermenting Gram-negative bacillus  33  50.8% 
Pseudomonas spp. (not P. aeruginosa19  – 
Acinetobacter baumannii  – 
Acinetobacter spp.  – 
S. maltophilia  – 
Otherb  – 
Enterobacteria  14  21.5% 
Enterobacter spp.  – 
K. pneumoniae  – 
P. mirabilis  – 
S. marcescens  – 
Otherc  – 
Gram-positive flora  11  16.9% 
Plasma-coagulase negative  – 
Staphylococcus  – 
Streptococcus spp.  – 
S. aureus Bacillus spp.  – 
Filamentous fungus (Penicillium)  3.1% 
Yeast  6.2% 
Candida spp.  – 
Rhodotorula  – 
Atypical mycobacteria (M. mucogenicum)  1.5% 
Total microorganisms isolated  65  – 
a

On 16 devices the culture was polymicrobial.

b

Roseomonas spp., Shewanella spp., Sphingomonas spp., Ochrobactrum spp.

c

Aeromonas spp., Pantoea spp.

Despite the microbial findings, the microorganism isolated in the nebuliser only corresponded with the isolate in respiratory samples in one of the 71 patients studied (1.4%). In this case, Proteus mirabilis was isolated in two sputum samples 14 and 24 days respectively after being found in the nebuliser.

There have been few studies similar to the one we present here, where we estimate the prevalence of microbial contamination in amphotericin B nebulising systems used by lung transplant patients. In a 2005 study4 patient adherence to the cleaning and disinfection protocol was poor, at only 39%. After those findings, patient education on care of the nebulising equipment was intensified.

The aforementioned Microbiological Procedure highlights the magnitude of the problem of nosocomial infection associated with semi-critical devices and the insufficient surveillance to which they are subjected. The procedure also strongly recommends periodic microbiological cultures as part of the quality control of the cleaning and disinfection process. However, this document does not review home-administered nebulised therapy devices in particular clinical contexts, such as cystic fibrosis or lung transplantation, which also require the same considerations.

Despite the limitations of our study, we believe that this work demonstrates the need to standardise appropriate mechanisms for the disinfection of nebuliser devices. We also consider it necessary to protocolise stringent microbiological monitoring of these devices to assess the process; especially in lung transplant and other immunologically compromised patients.

References
[1]
M. Peghin, V. Monforte, M.T. Martin-Gomez, I. Ruiz-Camps, C. Berasategui, B. Saez, et al.
10 years of prophylaxis with nebulized liposomal amphotericin B and the changing epidemiology of Aspergillus spp. infection in lung transplantation.
Transpl Int, 29 (2016), pp. 51-62
[2]
J.A. Koss, T.A. Conine, H.E. Eitzen, A.M. LoSasso.
Bacterial contamination potential of sterile, prefilled humidifiers and nebulizer reservoirs.
Heart Lung, 8 (1979), pp. 1117-1121
[3]
H. Blau, H. Mussaffi, M. Mei Zahav, D. Prais, M. Livne, B.M. Czitron, et al.
Microbial contamination of nebulizers in the home treatment of cystic fibrosis.
Child Care Health Dev, 33 (2007), pp. 491-495
[4]
V. Monforte, A. Román, J. Gavaldà, C. Bravo, V. Rodriguez, A. Ferrer, et al.
Contamination of the nebulization systems used in the prophylaxis with amphotericin B nebulized in lung transplantation.
Transplant Proc, 37 (2005), pp. 4056-4058
[5]
C.A. O’Malley.
Device cleaning and infection control in aerosol therapy.
Respir Care, 60 (2015), pp. 917-927

Please cite this article as: Fernández-Huerta M, Escobar R, Monforte V, Rodriguez V. Contaminación microbiana de dispositivos de nebulización de anfotericina B liposomal en pacientes trasplantados pulmonares. Enferm Infecc Microbiol Clin. 2019;37:211–212.

Copyright © 2018. Elsevier España, S.L.U. and Sociedad Española de Enfermedades Infecciosas y Microbiología Clínica
Article options
Tools
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos