Elsevier

Journal of Vascular Surgery

Volume 39, Issue 2, February 2004, Pages 279-287
Journal of Vascular Surgery

Clinical research study from the American Association for Vascular Surgery
Endovascular management of juxtarenal aneurysms with fenestrated endovascular grafting

Presented at the Fifty-first Annual Meeting of the American Association for Vascular Surgery, Chicago, Ill, Jun 8-11, 2003.
https://doi.org/10.1016/j.jvs.2003.09.050Get rights and content
Under an Elsevier user license
open archive

Abstract

Purpose

To evaluate the technical feasibility and short-term results of juxtarenal aneurysm repair with an endovascular graft that incorporated the visceral aortic segment with graft material.

Methods

Patients were studied prospectively after the implantation of an endovascular device with graft material extending proximal to the renal arteries, variably incorporating the superior mesenteric and celiac arteries. All patients were deemed to be high risk with respect to open surgical repair and had compromised proximal neck anatomy. Proximal neck lengths were ≤10 mm, or ≤15 mm with a challenging morphology (funnel shape or extensive thrombus). Fenestrations within the graft material were customized to accommodate visceral and renal vessels on the basis of computerized tomography (CT), angiography, or intravascular ultrasound data. Selected visceral ostia were protected with balloon-expandable stents after partial endograft deployment. All patients were evaluated with CT and kidney, ureters, and bladder x-ray at discharge and at 1, 6, and 12 months. Visceral duplex scan studies were performed at 1, 6, and 12 months.

Results

A total of 22 patients were enrolled in the study. Sixteen patients had short proximal necks (3-10 mm), and six had compromised necks of 10 to 15 mm in length. Endograft design included bifurcated (20) and tube (2) systems. All prostheses were implanted successfully without the acute loss of any visceral arteries. A total of 58 visceral vessels were incorporated (mean, 2.6 per patient) and most commonly included both renal arteries and the superior mesenteric artery. The mean follow-up was 6 months. There were no deaths within 30 days and no aneurysm-related deaths during the follow-up period. Two early (<30 days) and two late secondary interventions were performed, inclusive of two visceral artery stenoses detected with duplex scanning. The 30-day endoleak rate was 4.5%. The aneurysm sac decreased greater than 5 mm in 53 % of patients at 6 months and three of four patients at 12 months. Three patients developed renal insufficiency, only one of which required temporary hemodialysis.

Conclusions

The placement of an endovascular prosthesis with graft material that incorporates the visceral arteries is technically feasible. The occurrence of endoleaks appears to be relatively low. The increased sealing and fixation zones in this patient population should limit the late development of proximal endoleak or migration; however, this situation will require more patients and extended follow-up.

Cited by (0)

Competition of interest: Dr. Greenberg receives (or has received) research funding from the following companies: Boston Scientific, Cook Inc, Guidant, Medtronics, Sulzer-Vascutek, and W.L. Gore. Dr. Greenberg receives (or has received) consultant fees for speaking engagements or for services rendered from Boston Scientific and Cook.