Performance of a new interference-resistant glucose meter

https://doi.org/10.1016/j.clinbiochem.2009.09.010Get rights and content

Abstract

Objectives

Glucose meters are widely used in self and hospital monitoring of blood glucose. We examined the analytical performance of a StatStrip glucose monitoring system.

Design and methods

Linearity, % recovery and within-run imprecision were studied using glucose-spiked whole blood. A total of 120 heparinized samples were used in method comparison using a plasma hexokinase on the Dimension RxL MAX analyzer as the comparison method. Common interferences were tested on the StatStrip, Accu-Chek Advantage and the MediSense Optium glucose meters at low, middle and high glucose levels.

Results

The StatStrip assay showed excellent linearity and recovery. The coefficient of variations for imprecision were < 5%. This meter correlated well with the comparison method (y = 0.994X + 0.03; r = 0.995, Sy/x = 0.05 mmol/L, bias =  0.01 mmol/L). Of the three meters tested, only the StatStrip showed interference < 10% for all spiked levels of acetaminophen, ascorbic acid, maltose and hematocrit at three levels of glucose tested.

Conclusions

The StatStrip meter showed good performance and is suitable for point-of-care hospital glucose testing.

Introduction

It is well established that diabetes mellitus (DM) and diabetes-related complications are a major health concern worldwide. The American Diabetes Association (ADA) clinical practice guidelines recommend that self-monitoring of blood glucose (SMBG) is important in the management of type 1 and type 2 diabetes to decrease the risk of complications [1], [2]. In recent years the value of glucose meters in point-of-care testing (POCT) by medical professionals and SMBG by patients have been widely accepted. These modern devices are commercially available, easy to use, allow rapid turnaround times and quantify measurements in very small volumes of specimen [3], [4]. Meters can provide a whole blood glucose measurement based on glucose oxidase or glucose dehydrogenase methods and measure either amperometric changes based on the oxidation or dehydrogenation of glucose, or colorimetric changes by linking the product of the enzymatic reaction to a chromophore. The methodologies used and the specifications of these meters have been reviewed and discussed [5], [6], [7], [8].

Since glucose meters are important tools in SMBG, it is recommended that the meters be accurate, precise and reliable. Different analytical goals have been suggested for the performance of glucose meters. The current ADA recommendation is for a total error of < 5% at glucose concentrations of 1.7-22.2 mmol/L [9]. The new Clinical and Laboratory Standards Institute (CLSI) guideline states that the results from the POC glucose monitoring system should agree within ± 0.8 mmol/L of the laboratory analyzer values at glucose concentrations below 4.2 mmol/L and within ± 20 % of the laboratory analyzer value at glucose concentrations at or above 4.2 mmol/L [10]. Kost et al. [11] have proposed a modification of the previous CLSI guideline that the discrepancy between meters and routine laboratory analyzers should not exceed + 0.8 mmol/L for glucose values < 5.6 mmol/L and + 15% for glucose values > 5.6 mmol/L, and have suggested that 95% of the individual results should fall within these limits. According to International Organization for Standardization (ISO) 15197:2003, 95% of the individual glucose results must be within ± 0.8 mmol/L of the comparison-method results at glucose concentrations below 4.20 mmol/L and within ± 20% of the comparison-method results for glucose concentrations at or above 4.20 mmol/L [12]. Clarke Error Grid analysis can also be used to evaluate glucose meter performance [13]. The Food and Drug Administration (FDA) requires all meters to have an error of < 20% at blood glucose concentrations between 1.7-22.2 mmol/L [14]. The analytical performances of most glucose meters presently available have been extensively evaluated. Several authors have reported good correlation between various meters and routine laboratory analyzers [15], [16], [17], [18]. Our aims in this study were to evaluate the analytical performance of the Nova StatStrip glucose meter and to compare the results to a routine laboratory method. The effects of acetaminophen, ascorbic acid, maltose and hematocrit on three glucose strip technologies were also assessed.

Section snippets

Equipment

The following blood glucose meters were used in the study: the Nova StatStrip glucose meter (Nova Biomedical, Waltham, MA, USA), the Roche Accu-Chek Advantage (Roche Diagnostics, Indianapolis, IN, USA), and the MediSense Optium (Abbott Diagnostics, Abbott Park, IL, USA). The Nova StatStrip glucose meter uses a modified glucose oxidase (MGO) based amperometric test system with hematocrit and other interference correction, while the Roche Accu-Chek Advantage and the Abbott Optium use glucose

Linearity and recovery study

Linearity of the Nova StatStrip glucose meter was determined using spiked whole blood specimens. The meter showed good linearity of response within the range of 1.5–33.3 mmol/L. The calculated linear regression equation was: y = 1.003x + 0.07, r = 1.000 (Fig. 1). The means of the percent recovery at low, middle and high glucose concentrations for the StatStrip, Accu-Chek and Optium were as follows: 98.8%, 101.9%, and 98.8%; 112.5%, 82.6%, and 74.7%; 66.7%, 70.4%, and 75.8%, respectively. Only the

Discussion

Self-monitoring of blood glucose using portable glucose meters is important in diabetes control [1]. Major problems for these devices are bias, interference by hematocrit and various blood constituents, and very high protein concentrations. Agreement with the routine method in the central laboratory is particularly important if different glucose meters are used. The reliability of glucose meters has recently been improved and several studies have evaluated the use of blood glucose meters and

Acknowledgments

The authors thank the Nova Biomedical Corporation for supplying the StatStrip Blood Glucose Monitoring system.

References (40)

  • SlussP.M.

    Bedside blood glucose testing systems: current technology

    Point Care

    (2006)
  • American Diabetes Association

    Self-monitoring of diabetes

    Diabetes Care

    (1996)
  • Point-of-Care Blood-Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline

    (2002)
  • KostG.J. et al.

    Multicenter study of oxygen insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada

    Crit. Care Med.

    (1998)
  • Sinternational Organization for Standardization

    In vitro Diagnostic test systems—requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2003 (E)

    (2003)
  • ClarkeW.L. et al.

    Evaluating clinical accuracy of systems for self-monitoring of blood glucose

    Diabetes Care

    (1987)
  • BriggsA.L. et al.

    Self-monitoring blood glucose (SMBG): now and the future

    J. Pharm. Pract.

    (2004)
  • KilpatrickE.S. et al.

    Variations in sample pH and pO2 affect ExacTech meter glucose measurements

    Diabet. Med.

    (1994)
  • KostG.J. et al.

    Multicentre study of oxygen-insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada

    Crit. Care Med.

    (1998)
  • ChanceJ.J. et al.

    Technical evaluation of five glucose meters with data management capabilities

    Am. J. Clin. Pathol.

    (1999)
  • Cited by (27)

    • Determination of lactose in milk and milk-derived ingredients using biosensor-based techniques

      2023, Enzymes Beyond Traditional Applications in Dairy Science and Technology
    • Technical note: Glucose concentration in dairy cows measured using 6 handheld meters designed for human use

      2019, Journal of Dairy Science
      Citation Excerpt :

      All handheld glucose meters evaluated in the present study were designed for human use and intended as monitoring tools for diabetic patients. In human blood specimens, hematocrit interferes with glucose concentration readings (Vanavanan et al., 2010). However, the hematocrit of cows is lower than that of humans; therefore, evaluating glucose concentration with handheld meters designed for humans may result in an overestimation of glucose concentration (Vanavanan et al., 2010).

    • Clinical impact of improved point-of-care glucose monitoring in neonatal intensive care using Nova StatStrip: Evidence for improved accuracy, better sensitivity, and reduced test utilization

      2016, Clinical Biochemistry
      Citation Excerpt :

      It uses multi-well technology in combination with a modified glucose oxidase amperometric test system to autocorrect for hematocrit and background noise [11]. We and others have demonstrated its improved analytical performance, in particular, resistance to hematocrit interferences, excellent accuracy against laboratory plasma measurements, and excellent precision < 4% coefficient of variation (CV) [10–14]. However, these studies have focused primarily on blood sampling across the normal range of glucose homeostasis and were performed mostly by trained laboratory operators.

    • Feasibility of continuous glucose monitoring in critically Ill emergency department patients

      2012, Journal of Emergency Medicine
      Citation Excerpt :

      In this study, we used a StatStrip® Xpress™ glucose monitor (Nova Biomedical) as capillary glucose measuring reference. The ability of this monitor to eliminate the aforementioned interferences has been validated in previous studies (17–19). This study has several limitations.

    • Glycemic control in the ICU

      2011, Chest
      Citation Excerpt :

      In anemic patients, this correction results in POC devices reporting falsely high blood glucose concentrations,25,76 which may mean that hypoglycemia goes undetected, or inappropriate doses of insulin are administered. Some newer POC devices measure and correct for hematocrit77; thus, it is important that clinicians understand the characteristics of the particular POC device used in their own ICU. Blood glucose measurements using glucose oxidase systems are prone to errors caused by oxygen effects.

    View all citing articles on Scopus
    View full text