Original article
Reproducibility and Repeatability of Central Corneal Thickness Measurement in Keratoconus Using the Rotating Scheimpflug Camera and Ultrasound Pachymetry

https://doi.org/10.1016/j.ajo.2007.07.021Get rights and content

Purpose

To assess repeatability, reproducibility, and agreement of rotating Scheimpflug camera (Pentacam Oculus, Wetzlar, Germany) and ultrasound pachymetry in measuring central thickness of keratoconic corneas.

Design

Method-comparison study.

Methods

In 33 patients with keratoconus (one eye per patient), two examiners each used both pachymetric methods to measure central corneal thickness (CCT); in the same session, measurements then were repeated by examiner 1 (A.M.). The difference between two examiners, and between first and second measurements by examiner 1, with both methods and the difference between the two pachymetric methods in measuring central thickness of keratoconic corneas were noted.

Results

With the rotating Scheimpflug camera, interexaminer correlation was higher (intra-class correlation coefficient [ICC], 0.98 vs 0.76) and inter-examiner variability was lower (95% limits of agreement [95% LoA], −14.8 to 13.8 μm vs −18.0 to +49.5 μm) than with ultrasound pachymetry. Both methods showed close first- to second-measurement correlation (ICC, > 90), but the rotating Scheimpflug camera had lower variability (95% LoA, −14.5 to 14.2 μm vs −27.4 to 26.0 μm). Mean CCT was 478.9 ± 34.6 μm with the rotating Scheimpflug camera and 486.6 ± 30 μm with ultrasound pachymetry. Although the mean difference was small (−7.8 μm), the 95% LoA (−43.8 to 28.2 μm) showed that the difference between the two methods can be considerable.

Conclusions

In keratoconic corneas, the rotating Scheimpflug camera provides measurements of central thickness that are more reproducible and repeatable than those obtained with ultrasound pachymetry. The rotating Scheimpflug camera seems to be suitable for disease staging and follow-up, when corneal thickness measurements may be repeated over time by different examiners.

Section snippets

Methods

CCT was measured with the rotating Scheimpflug camera and ultrasound pachymetry in 33 patients with previously diagnosed keratoconus (22 men, 11 women; mean age ± standard deviation [SD], 37 ± six years; range, 19 to 61 years). The initial diagnosis of keratoconus was based on clinical slit-lamp findings and associated characteristic Placido-based topographic patterns. Slit-lamp findings included one or more of the following signs: stromal thinning, conical protrusion, Fleischer ring, Vogt

Reproducibility and Repeatability of the Two Methods in Central Thickness Measurement of Keratoconic Corneas

Analysis of inter-examiner reproducibility revealed a lower ICC, higher coefficient of repeatability, and wider 95% LoA for ultrasound pachymetry compared with the rotating Scheimpflug camera (Table 1). The 95% LoA for ultrasound pachymetry indicated that differences between examiners could be of clinical significance with this method. Figure 1 graphically illustrates the difference between the two examiners in measuring each individual CCT, with both methods, through Bland-Altman plots.

Discussion

In Keratoconus, the cornea tends to be thinner than normal, both centrally and peripherally,25, 26 so that reproducible, repeatable, and accurate measurements of corneal thickness are required for the diagnosis, staging, and follow-up of this disease, as well as for planning surgical procedures. Although a comprehensive clinical examination plus Placido-based topography have good accuracy in detecting keratoconus, measurement of corneal thickness may provide useful corroborative evidence for

Ugo de Sanctis, MD, PhD, completed the post-graduate training in Ophthalmology, and the post-graduation doctorate in Cornea and External Diseases at the Ophthalmology Institute, University of Turin, Italy. Currently, Dr de Sanctis is an Assistant Professor of Ophthalmology and Head of Cornea Service at the Medical University of Turin. His clinical and research interests include corneal and external diseases, refractive surgery, and cataract surgery.

References (32)

  • S. Amano et al.

    Comparison of central corneal thickness measurement by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography

    Ophthalmology

    (2006)
  • U. de Sanctis et al.

    Graft central thickness measurement by rotating Scheimpflug camera and ultrasound pachymetry after penetrating keratoplasty

    Ophthalmology

    (2007)
  • O.O. Ucakhan et al.

    Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry

    J Cataract Refract Surg

    (2006)
  • M.B. Shapiro et al.

    Anterior clear spaces in keratoconus

    Ophthalmology

    (1986)
  • E.E. Manche et al.

    Deep lamellar keratoplasty using viscoelastic dissection

    Arch Ophthalmol

    (1999)
  • G.R. Melles et al.

    A quick surgical technique for deep, anterior lamellar keratoplasty using visco-dissection

    Cornea

    (2000)
  • Cited by (136)

    View all citing articles on Scopus

    Ugo de Sanctis, MD, PhD, completed the post-graduate training in Ophthalmology, and the post-graduation doctorate in Cornea and External Diseases at the Ophthalmology Institute, University of Turin, Italy. Currently, Dr de Sanctis is an Assistant Professor of Ophthalmology and Head of Cornea Service at the Medical University of Turin. His clinical and research interests include corneal and external diseases, refractive surgery, and cataract surgery.

    View full text