Calibración de PET/TC para adquisición de estudio de cuerpo estándar postratamiento con esferas de resina de ytrio-90
PET/CT calibration for post-treatment standard body scan acquisition with yttrium-90 Resin Micro-SPHERES
Visitas
64
J.A. Uña-Gorospea,
Autor para correspondencia
junagor@gobiernodecanarias.org

Autor para correspondencia.
, C. Baso-Pérezb, C. Cárdenas-Negroc, J.A. Merino-Gestosod, B.C. Portas-Ferradáse, A. Allende-Rieraf
a Nuclear Medicine Department, Hospital Universitario Nuestra Señora de Candelaria, Santa Cruz de Tenerife, Islas Canarias, España
b Radiology Department, Hospital Universitario de Canarias, Santa Cruz de Tenerife, Islas Canarias, España
c Nuclear Medicine Department, Hospital Universitario Nuestra Señora de Candelaria, Santa Cruz de Tenerife, Islas Canarias, España
d Physicist, Medical Physics Department, Hospital Universitario Nuestra Señora de Candelaria, Santa Cruz de Tenerife, Islas Canarias, España
e Physicist, Medical Physics Department, Hospital Universitario Nuestra Señora de Candelaria, Santa Cruz de Tenerife, Islas Canarias, España
f Nuclear Medicine Department, Hospital Universitario Nuestra Señora de Candelaria, Santa Cruz de Tenerife, Islas Canarias, España
Este artículo ha recibido
Recibido 09 noviembre 2021. Aceptado 22 mayo 2022
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (5)
Mostrar másMostrar menos
Resumen
Introducción

Entre los objetivos en los tratamientos con esferas marcadas con ytrio-90 figura demostrar que se consigue la dosis tumoricida y que se evitan dosis hepatotóxicas, así como que no existe deposición extrahepática.

Material y métodos

Utilizamos diferentes cantidades de ytrio y un maniquí NEMA NU2-2007 para comprobar si el tomógrafo mantiene una respuesta que permita hacer cálculos dosimétricos reales.

Resultados

El tomógrafo Gemini responde de manera lineal en un rango amplio de actividades con una R2=0,9983.

Conclusión

La capacidad de detección del equipo PET nos permite realizar estudios de cuerpo estándar con finalidad dosimétrica con tiempos de 5min por BED, que además hacen posible verificar la ausencia de deposición extrahepática de cantidades significativas de esferas de ytrio-90.

Palabras clave:
PET
Esferas de resina
Ytrio-90
SIR-Spheres®
Calibración
Abstract
Introduction

Among the objectives in treatments with yttrium-90 spheres is to demonstrate that the tumoricidal dose is achieved and that hepatotoxic doses are avoided, as well as that there is no extrahepatic deposition.

Material and methods

We use different amounts of yttrium-90 resin micro-spheres and a NEMA NU2-2007 phantom to check if the scanner maintains a response that allows real dosimetric calculations.

Results

The Gemini tomograph responds linearly in a wide range of activities with R2=0.9983.

Conclusion

The detection capacity of the PET equipment allows us to carry out standard body studies with dosimetric purposes with times of five minutes per BED. It also make possible to verify the absence of extrahepatic deposition of significant amounts of yttrium 90 spheres.

Keywords:
PET
Resin spheres
Yttrium-90
SIR-Spheres®
Calibration

Artículo

Opciones para acceder a los textos completos de la publicación Revista Española de Medicina Nuclear e Imagen Molecular
Socio
Si es usted socio de la Sociedad Española de Medicina Nuclear e Imagen Molecular (SEMNIM) puede acceder al texto completo de los contenidos de la Revista Española de Medicina Nuclear e Imagen Molecular desde los enlaces a la revista publicados en la web de la SEMNIN (enlace a https://semnim.es/iniciar-sesion/), previo inicio de sesión como socio. Si tiene problemas de acceso puede contactar con la Secretaría Técnica de la SEMNIM en el correo electrónico secretaria.tecnica@semnim.es o en el teléfono: + 34 619 594 780
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Revista Española de Medicina Nuclear e Imagen Molecular

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
Herramientas
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos