Automated liquid media have represented a major advance in the culture of mycobacteria, limiting the use of solid media and even removing them. In this study, we investigated whether the Bactec MGIT automated system fails to detect mycobacteria and their possible causes. Over a period of three years (2023–2025), a total of 9.111 cultures for mycobacteria were performed using automated liquid media (Bactec MGIT). All negative tubes were inspected visually, and those with suspected growth were stained with Ziehl-Neelsen. Those that were positive were reinoculated, and the grown colonies were identified by molecular detection (Xpert MTB/RIF) and proteomics (MALDI-TOF). A total of 0.2% of all tubes incubated in this system were considered a detection failure, representing 11.1% of all mycobacterial isolates. A total of 22 cases were recorded, of which 8 (36.3%) were identified as M. tuberculosis complex (MTC) and the remaining 14 as NTMs (63.6%): M. lentiflavum, M. gordonae, M. avium, M. abscessus subsp. abscessus, M. chelonae, M. mageritense, M. celatum, and M. paragordonae. In 6 out of 8 patients who were undergoing anti-tuberculosis treatment an MTC species was isolated. When an MNT was isolated, only those isolated in non-respiratory samples were considered clinically significant. A final visual inspection of these negative tubes in the absence of molecular techniques, would be the only way to detect these cases.
Los medios líquidos automatizados han supuesto un gran avance en el cultivo de micobacterias, reduciendo el tiempo de crecimiento y facilitando la identificación. Han reducido el empleo de medios sólidos, siendo en muchas ocasiones el único sistema de cultivo. En este estudio hemos investigado si el sistema automático BACTEC MGIT falla en la detección, generando fallos de detección y cuáles son sus posibles causas. Durante un periodo de 3 años (2023–2025) se han realizado en nuestro centro un total de 9.111 cultivo para micobacterias en medios líquidos automatizados (BACTEC MGIT). Los tubos negativos se inspeccionaron visualmente y aquellos con sospecha de crecimiento se tiñeron con Ziehl-Neelsen. Los que fueron positivos se resembraron y las colonias se identificaron por detección molecular (Xpert MTB/RIF) y proteómica (MALDI-TOF). Un 0,2% de todos los tubos incubados en este sistema presentaron fallos de detección, lo que supone un 11,1% de todos los aislamientos de micobacterias. En total se registraron 22 casos, de los cuales 8 (36,3%) se identificaron como M.tuberculosis complex (MTC) y los 14 restantes como MNTs (63,6%): M.lentiflavum, M.gordonae,M.avium, M.abscessus subsp. abscessus, M.chelonae, M.mageritense, M.celatum y M.paragordonae. Cuando se aisló una especie del MTC (en 6 de 8 pacientes), éstos estaban siguiendo tratamiento antituberculoso, uno de los cuáles con baciloscopia positiva. Cuando se aisló una MNT, únicamente se dio importancia clínica a las aisladas en muestras no respiratorias. Una última inspección visual de estos tubos negativos y en ausencia de técnicas moleculares, sería la única forma de detectar estos casos.